[Turbo(fast) spin echo at 0.5 T: effect of echo distance and echo number on image contrast].

Vahlensieck M, Seelos K, Gieseke J, Reiser M

Аннотация на английскомСтатьяПеревод в очереди

Аннотация

Signal intensities, signal-to-noise ratios (S/N), and contrast-to-noise ratios (C/N) in dependence on the echo distance and echo number of the new turbo (fast) spin-echo technique are analysed by phantom, volunteer and patient measurements. With increasing echo number (3 to 15 echoes), signal intensities increase (25 to 50%), S/N varies between different tissues, and C/N of fat to water decreases (70%). A high echo number, although it shortens the imaging time, is thus not always the best choice since the fat/water contrast is lost. However, a high fat/water contrast is advantageous for several MR applications.

Для цитирования: Vahlensieck M, Seelos K, Gieseke J, Reiser M. [Turbo(fast) spin echo at 0.5 T: effect of echo distance and echo number on image contrast].. RoFo : Fortschritte auf dem Gebiete der Rontgenstrahlen und der Nuklearmedizin. 1993. DOI: 10.1055/s-2008-1032644.