MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1148/radiology.200.2.8685359

Медицинская библиотека · Radiology · 1996

Breath-hold, contrast-enhanced, three-dimensional MR angiography.

Leung D A, McKinnon G C, Davis C P, Pfammatter T, Krestin G P, Debatin J F

СтатьяПеревод в очереди

Аннотация

Breath-hold gadolinium-enhanced three-dimensional magnetic resonance (MR) angiography was performed with a high-performance gradient MR imaging system in 10 volunteers with no history of vascular abnormalities (breath-hold interval, 28 seconds; repetition time, 4.0 msec; echo time, 1.9 msec; 0.4 mmol gadoterate meglumin per kilogram of body weight). High-resolution three-dimensional time-of-flight angiograms of the aorta, renal arteries, pulmonary arteries, pelvic arteries, and portal venous system were successfully acquired in all cases.

Для цитирования: Leung D A, McKinnon G C, Davis C P, Pfammatter T, Krestin G P, Debatin J F. Breath-hold, contrast-enhanced, three-dimensional MR angiography.. Radiology. 1996. DOI: 10.1148/radiology.200.2.8685359.