DOI 10.1128/cdli.12.10.1231-1234.2005

Медицинская библиотека · Clinical and diagnostic laboratory immunology · 2005

Evaluation of LIAISON Treponema Screen, a novel recombinant antigen-based chemiluminescence immunoassay for laboratory diagnosis of syphilis.

Marangoni Antonella, Sambri Vittorio, Accardo Silvia, Cavrini Francesca, D'Antuono Antonietta, Moroni Alessandra, Storni Elisa, Cevenini Roberto

Сравнительное исследованиеОценочное исследованиеСтатьяПеревод в очереди

Аннотация

The purpose of this study was to evaluate the diagnostic performance of LIAISON Treponema Screen (DiaSorin, Saluggia, Italy), a new automated chemiluminescence immunoassay (CLIA), in comparison with that of rapid plasma reagin (RPR) and the following currently used treponemal tests: hemagglutination test (TPHA), immunoenzymatic assay (EIA), and Western blot (WB). First, a retrospective study was performed with a panel of 2,494 blood donor sera, a panel of 131 clinical and serologically characterized syphilitic sera, and 96 samples obtained from subjects with potentially interfering diseases or conditions. A prospective study was also performed by testing 1,800 unselected samples submitted to the Microbiology Laboratory of the St. Orsola Hospital in Bologna, Italy, for routine screening for syphilis. As expected, RPR was the least specific method, especially when potentially cross-reacting sera were tested. On the contrary, all of the treponemal tests proved to be very specific (99.9%) and they performed with the following sensitivities: 100% (WB), 99.2% (CLIA), 95.4% (EIA), and 94.7% (TPHA).

Для цитирования: Marangoni Antonella, Sambri Vittorio, Accardo Silvia, Cavrini Francesca, D'Antuono Antonietta, Moroni Alessandra, Storni Elisa, Cevenini Roberto. Evaluation of LIAISON Treponema Screen, a novel recombinant antigen-based chemiluminescence immunoassay for laboratory diagnosis of syphilis.. Clinical and diagnostic laboratory immunology. 2005. DOI: 10.1128/cdli.12.10.1231-1234.2005.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.