Медицинская библиотека · Alzheimer's & dementia : the journal of the Alzheimer's Association · 2026

Analytical and clinical validation of the Lumipulse G plasma p-tau217 assay for clinical implementation.

Arslan Burak, Gobom Johan, Andreasson Ulf, Montoliu-Gaya Laia, Benedet Andrea L, Dittrich Anna, Kern Silke, Skoog Ingmar, Ashton Nicholas J, Chan Tevy, Rahmouni Nesrine, Rosa-Neto Pedro, Blennow Kaj, Zetterberg Henrik, Kvartsberg Hlin

СтатьяВалидационное исследованиеПеревод в очереди

Аннотация

INTRODUCTION

Although several plasma tau phosphorylated at threonine 217 (p-tau217) immunoassays are now available, comprehensive analytical validation remains limited. We therefore evaluated the Lumipulse G plasma p-tau217 assay and verified established cutoffs for clinical implementation.

METHODS

This study was conducted in two phases: (1) analytical validation of the Lumipulse G plasma p-tau217 assay, assessing precision, lower limit of quantification (LLoQ), selectivity, stability, and interference; and (2) verification of previously established cutoffs using a two-threshold approach in 37 samples with confirmed cerebrospinal fluid (CSF) Aβ42/Aβ40 status.

RESULTS

The assay demonstrated strong analytical performance, with repeatability and intermediate precision (%CV < 7%) and an LLoQ of 0.12 pg/mL. The p-tau217 was stable across freeze-thaw cycles but less so at 4°C, and hemolysis > 2% introduced variability. Cutoff verification showed 97% reproducibility with excellent agreement (ρ = 0.99, p < 0.0001).

DISCUSSION: The Lumipulse assay showed robust analytical performance and reproducibility, supporting clinical use, though certified reference materials are still needed for standardization.

Для цитирования: Arslan Burak, Gobom Johan, Andreasson Ulf, Montoliu-Gaya Laia, Benedet Andrea L, Dittrich Anna, Kern Silke, Skoog Ingmar, Ashton Nicholas J, Chan Tevy, Rahmouni Nesrine, Rosa-Neto Pedro, Blennow Kaj, Zetterberg Henrik, Kvartsberg Hlin. Analytical and clinical validation of the Lumipulse G plasma p-tau217 assay for clinical implementation.. Alzheimer's & dementia : the journal of the Alzheimer's Association. 2026. DOI: 10.1002/alz.71374.