Analytical and clinical validation of the Lumipulse G plasma p-tau217 assay for clinical implementation.
Аннотация
INTRODUCTION
Although several plasma tau phosphorylated at threonine 217 (p-tau217) immunoassays are now available, comprehensive analytical validation remains limited. We therefore evaluated the Lumipulse G plasma p-tau217 assay and verified established cutoffs for clinical implementation.
METHODS
This study was conducted in two phases: (1) analytical validation of the Lumipulse G plasma p-tau217 assay, assessing precision, lower limit of quantification (LLoQ), selectivity, stability, and interference; and (2) verification of previously established cutoffs using a two-threshold approach in 37 samples with confirmed cerebrospinal fluid (CSF) Aβ42/Aβ40 status.
RESULTS
The assay demonstrated strong analytical performance, with repeatability and intermediate precision (%CV < 7%) and an LLoQ of 0.12 pg/mL. The p-tau217 was stable across freeze-thaw cycles but less so at 4°C, and hemolysis > 2% introduced variability. Cutoff verification showed 97% reproducibility with excellent agreement (ρ = 0.99, p < 0.0001).
DISCUSSION: The Lumipulse assay showed robust analytical performance and reproducibility, supporting clinical use, though certified reference materials are still needed for standardization.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.