DOI 10.1097/rli.0000000000001159

Медицинская библиотека · Investigative radiology · 2025

In Vivo Bone Mineral Density Assessment With Spectral Localizer Radiographs From Photon-Counting Detector CT: Prospective Comparison With DXA.

Moser Lukas Jakob, Klambauer Konstantin, Diaz Machicado Maria Carolina, Frey Diana, Mergen Victor, Eberhard Matthias, Nowak Tristan, Schmidt Bernhard, Flohr Thomas, Distler Oliver, Alkadhi Hatem

СтатьяСравнительное исследованиеПеревод в очереди

Аннотация

PURPOSE: The aim of this study was to determine in a prospective patient study the accuracy of areal bone mineral density (aBMD) measurements with spectral localizer radiographs obtained with a clinical photon-counting detector computed tomography (PCD-CT) scanner in comparison with dual-energy x-ray absorptiometry (DXA).

METHODS

In this institutional review board-approved, prospective study, 41 patients (15 females, 26 males; mean age 61.3 years, age range 35-78 years) underwent PCD-CT of the abdomen with a spectral localizer radiograph (tube voltage 140 kVp, tube current 30 mA) and DXA within a median of 45 days. aBMD values were derived for lumbar vertebrae L1-L4 from both methods and were compared with linear regression, Pearson correlation, intraclass correlation coefficients (ICCs), and Bland-Altman plots. T-scores were calculated on a patient level and were compared between methods.

RESULTS

DXA and spectral localizer radiographs showed strong correlation in aBMD measurements ( R = 0.97, P < 0.001) and patient level T-scores ( R = 0.99, P < 0.001). There was a strong agreement between aBMD from both methods (ICC, 0.96; 95% CI, 0.94-0.97). Bland-Altman analysis revealed a very small mean difference in aBMD between methods (mean absolute error 0.019 g/cm 2 ) with narrow limits of agreement (-0.083 g/cm 2 to 0.121 g/cm 2 ). Similarly, there were small differences in regard to the T-score (mean absolute error 0.156) with narrow limits of agreement (-0.422 to 0.734) between methods. ICCs indicated an excellent agreement between T-scores from DXA and spectral localizer radiographs (ICC, 0.98; 95% confidence interval, 0.95-0.99).

CONCLUSIONS

Our prospective patient study indicates that spectral localizer radiographs obtained with a clinical PCD-CT system enable accurate quantification of the lumbar bone areal mineral density. This opens up the opportunity for opportunistic screening of osteoporosis in patients who undergo CT for other indications.

Для цитирования: Moser Lukas Jakob, Klambauer Konstantin, Diaz Machicado Maria Carolina, Frey Diana, Mergen Victor, Eberhard Matthias, Nowak Tristan, Schmidt Bernhard, Flohr Thomas, Distler Oliver, Alkadhi Hatem. In Vivo Bone Mineral Density Assessment With Spectral Localizer Radiographs From Photon-Counting Detector CT: Prospective Comparison With DXA.. Investigative radiology. 2025. DOI: 10.1097/rli.0000000000001159.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.