DOI 10.5137/1019-5149.jtn.49191-25.2
Finite Element Analysis of Mechanical Stress Distribution in Titanium and PEEK Rod Constructs for Isthmic L5-S1 Spondylolisthesis.
Аннотация
AIM: To compare the biomechanical behavior of titanium and polyetheretherketone (PEEK) rods using finite element analysis (FEA) in a Grade I L5-S1 isthmic spondylolisthesis model.
MATERIAL AND
METHODS
An anatomically accurate 3D finite element model of the L3?S1 spine was created from CT data of a healthy 30-year-old male. A unilateral pars defect at L5 simulated Grade I spondylolisthesis. Instrumented models with titanium and PEEK rods were subjected to five physiological loading conditions. Von Mises stresses and displacement vectors were analyzed and stress distributions on bone, screws and rods were measured.
RESULTS
Titanium rods absorbed significantly higher loads across all loading conditions (e.g., 117 N vs. 34.5 N in compression; p < 0.01), while PEEK constructs allowed greater load sharing with the bone. Screws in titanium systems experienced higher stresses, particularly in posterior shear and axial rotation. Both materials effectively reduced stress at the pars defect.
CONCLUSION
Titanium constructs offer high rigidity but concentrate stress on implants, potentially increasing complication risk. PEEK rods demonstrated a more physiological load distribution, reducing implant stress without compromising stability. PEEK rods may serve as a biomechanically advantageous alternative to titanium in select spondylolisthesis cases. Further clinical validation is warranted.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.