Is Fretting Corrosion at the Modular Junction of Revision Total Knee Arthroplasty Actually a Problem?

Hall Deborah J, Wong John, Wright Jennifer L, Hochstatter Julia P, Burnett Robert A, Miller Amy, Levine Brett R, Pourzal Robin, Terhune E Bailey

СтатьяПеревод в очереди

Аннотация

BACKGROUND

Corrosion at modular junctions in total knee arthroplasty (TKA) implants is recognized, but its clinical relevance remains unclear. This study evaluated corrosion at modular stem-bore junctions and examined whether severity correlated with radiographic evidence of adverse local tissue reactions or osteolysis.

METHODS

We evaluated 196 retrieved modular stem and bore components from 98 revision TKA patients, including taper (n = 73) and threaded (n = 25) designs. The median time in situ was 24 months for taper and 15 months for threaded junctions (P = 0.052). Periprosthetic joint infection and aseptic loosening were the most common indications for component explantation. Fretting corrosion at the modular junction was graded using a modified Goldberg Score (mGS) and quantified using an optical coordinate measuring machine. Radiographs were evaluated. Statistical analyses included Mann-Whitney U-, Chi-Square, and Spearman's correlation tests.

RESULTS

Taper junctions demonstrated greater corrosion than threaded designs (P < 0.001). The median mGS for modular tapers was 4 for femoral components, 3 for femoral stems, and two for both tibial components and stems. Threaded junctions showed minimal damage across all interfaces with a median score of 1. Damage patterns included fretting wear, corrosion features, and dark deposits. Among cobalt-chromium (CoCr) femoral bore tapers with mGS of 4, median material loss was 0.56 mm3. Titanium alloy femoral and tibial stem tapers showed a median loss of 0.071 mm3, and 0.067 mm3, respectively. Corrosion severity did not correlate with localized radiographic findings or indication for revision.

CONCLUSIONS

In this study of modular TKA components, taper junctions exhibited significantly more corrosion than threaded junctions. However, corrosion severity was not associated with radiographic findings or etiology of failure. Corrosion products differed from those typically observed in THA, suggesting a distinct mechanism. Further research is required to clarify potential clinical implications.

Для цитирования: Hall Deborah J, Wong John, Wright Jennifer L, Hochstatter Julia P, Burnett Robert A, Miller Amy, Levine Brett R, Pourzal Robin, Terhune E Bailey. Is Fretting Corrosion at the Modular Junction of Revision Total Knee Arthroplasty Actually a Problem?. The Journal of arthroplasty. 2026. DOI: 10.1016/j.arth.2026.05.069.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.