Quantitative assessment of areal bone mineral density using multi-energy localizer radiographs from photon-counting detector CT.
СтатьяПоддержка исследования: вне правительства СШАНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Jasper Sören, Swicklik Joseph, Baffour Francis, Ferrero Andrea, El Sadaney Ahmed O, Shanblatt Elisabeth, Nowak Tristan, McCollough Cynthia, Rajendran Kishore
Аннотация
Objective.To assess the accuracy and stability of areal bone-mineral-density (aBMD) measurements from multi-energy CT localizer radiographs acquired using photon-counting detector (PCD) CT.Approach.A European Spine Phantom (ESP) with hydroxyapatite (HA 0.5, 1.0 and 1.5 g cm-2) was scanned using clinical PCD-CT and a dual-energy x-ray absorptiometry (DXA) to compare aBMD values. To assess aBMD stability and reproducibility, PCD-localizers were acquired twice a day for one week, and once per week for five weeks. Multiple ESP anteroposterior thicknesses (18, 26, 34, and 40 cm) were achieved using a synthetic gel layer and scanned across eight tube current values for both 120 kV (15-120 mA) and 140 kV (10-80 mA). One-way analysis of variance was performed for statistical significance (p< 0.05 considered significant). Quantitative HA and water maps were reconstructed using a prototype material-decomposition software, and aBMD was calculated after background correction.In vivoperformance of PCD-based aBMD was illustrated using a patient scan acquired at 140 kV and 40 mA, and lumbar aBMD values were compared with DXA.Main results.The ESP aBMD values from PCD-localizers demonstrated excellent day-to-day stability with a coefficient-of-variation ranging from 0.42% to 0.53%, with mean absolute percentage errors (MAPE) of less than 5% for all three vertebral bodies. The coefficient-of-variation for weekly scans ranged from 0.17% to 0.60%, again with MAPE below 5% for all three vertebral bodies. Across phantom sizes and tube currents, the MAPE values varied ranging from 1.84% to 13.78% for 120 kV, and 1.38%-9.11% for 140 kV. No significant difference was found between different tube currents. For the standard phantom size, DXA showed 11.21% MAPE whereas PCD-CT showed 3.04% MAPE. For the patient scan, deviation between PCD-based aBMD values and those obtained from DXA ranged from 0.07% to 9.82% for different lumbar vertebra.Significance.This study highlights the accuracy and stability of PCD-CT localizers for measuring aBMD. We demonstrated aBMD accuracy across different sizes and showed that higher radiation doses did not inherently increase aBMD accuracy.