Finite Element Analysis of Influence of Axial Position of Center of Rotation of a Cervical Total Disc Replacement on Biomechanical Parameters: Simulated 2-Level Replacement Based on a Validated Model.

Li Yang, Zhang Zhenjun, Liao Zhenhua, Mo Zhongjun, Liu Weiqiang

СтатьяПеревод в очереди

Аннотация

BACKGROUND

Finite element models have been widely used to predict biomechanical parameters of the cervical spine. Previous studies investigated the influence of position of rotational centers of prostheses on cervical biomechanical parameters after 1-level total disc replacement. The purpose of this study was to explore the effects of axial position of rotational centers of prostheses on cervical biomechanics after 2-level total disc replacement.

METHODS

A validated finite element model of C3-C7 segments and 2 prostheses, including the rotational center located at the superior endplate (SE) and inferior endplate (IE), was developed. Four total disc replacement models were used: 1) IE inserted at C4-C5 disc space and IE inserted at C5-C6 disc space (IE-IE), 2) IE-SE, 3) SE-IE, and 4) SE-SE. All models were subjected to displacement control combined with a 50 N follower load to simulate flexion and extension motions in the sagittal plane. For each case, biomechanical parameters, including predicted moments, range of rotation at each level, facet joint stress, and von Mises stress on the ultra-high-molecular-weight polyethylene core of the prostheses, were calculated.

RESULTS

The SE-IE model resulted in significantly lower stress at the cartilage level during extension and at the ultra-high-molecular-weight polyethylene cores when compared with the SE-SE construct and did not generate hypermotion at the C4-C5 level compared with the IE-SE and IE-IE constructs.

CONCLUSIONS

Based on the present analysis, the SE-IE construct is recommended for treating cervical disease at the C4-C6 level. This study may provide a useful model to inform clinical operations.

Для цитирования: Li Yang, Zhang Zhenjun, Liao Zhenhua, Mo Zhongjun, Liu Weiqiang. Finite Element Analysis of Influence of Axial Position of Center of Rotation of a Cervical Total Disc Replacement on Biomechanical Parameters: Simulated 2-Level Replacement Based on a Validated Model.. World neurosurgery. 2017. DOI: 10.1016/j.wneu.2017.07.079.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.