Polyester fibre prosthetic anterior cruciate ligament implant rupture: necrosis of ingrown connective tissue.
Клинический случайСтатьяНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Keen C E, Amis A A
Аннотация
AIMS: To describe the histopathological and microanalytical features in seven cases of ruptured Apex(R) polyester (Terylene(R)) fibre anterior cruciate ligament prosthesis.
METHODS AND RESULTS: Transmitted and polarized light microscopy was performed in all cases; one case was investigated by immunohistochemistry, transmission electron microscopy and scanning electron microscopy, with backscatter and X-ray detectors for elemental microanalysis. For comparison we also studied synovial biopsy material and unused polyester fibres. In the excised ligaments there was much ingrowth of fibrous tissue accompanying a florid giant cell reaction to the individual intact polyester fibres throughout the ligaments. Phagocytosis of particles of prosthesis-derived material was demonstrated and a striking finding was of necrosis of the ingrown connective tissue in the central portions of the ligaments. Hyalinized areas and 'neoligament growth' were less striking. A consistent finding in the polyester fibres was of small particles containing antimony, used as a catalyst in the manufacturing process.
CONCLUSIONS: The pattern of reaction to the prosthetic material and the presence of necrosis differ from previous descriptions in animal and human explants of this and other prosthesis types. The mechanical effect of the necrosis is unlikely to be of significance with this ligament, which is load-bearing ab initio.
METHODS AND RESULTS: Transmitted and polarized light microscopy was performed in all cases; one case was investigated by immunohistochemistry, transmission electron microscopy and scanning electron microscopy, with backscatter and X-ray detectors for elemental microanalysis. For comparison we also studied synovial biopsy material and unused polyester fibres. In the excised ligaments there was much ingrowth of fibrous tissue accompanying a florid giant cell reaction to the individual intact polyester fibres throughout the ligaments. Phagocytosis of particles of prosthesis-derived material was demonstrated and a striking finding was of necrosis of the ingrown connective tissue in the central portions of the ligaments. Hyalinized areas and 'neoligament growth' were less striking. A consistent finding in the polyester fibres was of small particles containing antimony, used as a catalyst in the manufacturing process.
CONCLUSIONS: The pattern of reaction to the prosthetic material and the presence of necrosis differ from previous descriptions in animal and human explants of this and other prosthesis types. The mechanical effect of the necrosis is unlikely to be of significance with this ligament, which is load-bearing ab initio.