Subcapital Femoral Neck Fracture Despite Cement-Augmented Cephalomedullary Nail Fixation for an Osteoporotic Intertrochanteric Fracture: A Case Report and Position- and Sliding-Based Decision Guide.

Yokoo Suguru, Okada Yukimasa, Ohno Kyotaro, Ichikawa Takahiko, Terada Chuji, Yamana Keiya

Клинический случайСтатьяПеревод в очереди

Аннотация

Background/

Objectives

Cement augmentation of cephalomedullary head elements can improve the purchase of osteoporotic bone; however, it does not eliminate the need for accurate implant positioning or the preservation of sliding. We report the case of an 87-year-old woman who underwent intramedullary nailing with a cement-augmented helical blade for intertrochanteric fracture.

Methods

This is a single-patient case report. Calibrated radiographic measurements-tip-apex distance (TAD), calcar-referenced TAD (CalTAD), neck-shaft angle (NSA), and telescoping-were obtained immediately postoperatively and at 4, 7, 12, and 15 months. CT was performed at postoperative week 1 and at failure, and MRI was performed for clinical deterioration. In addition, a targeted narrative review summarizes the evidence on the head-element position, sliding behavior, reduction alignment, and augmentation.

Results

Immediate postoperative indices were within the accepted targets: TAD 22.6 mm, CalTAD 22.8 mm, NSA 134°, with the head element inferior on the anteroposterior view and central on the lateral view. Rehabilitation proceeded with full weight bearing as tolerated. Early telescoping was minimal (3.8-3.9 mm). Between 7 and 15 months, progressive varus with shortening of TAD/CalTAD and little additional telescoping was observed, radiographically consistent with relative proximal migration of the head-cement complex and a cleavage plane along the inferior cement mantle, culminating in a subcapital femoral neck fracture with the implant in situ. Emphasis should be placed on accurate implant positioning and preservation of sliding capacity, because cement augmentation alone may not prevent mechanical failure when the implant position or load transfer is suboptimal.

Conclusions

Cement augmentation stiffens the interface and reduces micromotion but does not neutralize malposition-induced stresses. Accurate positioning, preservation of sliding, and timely conversion when sliding fails to progress are advisable; these findings are hypothesis-generating from a single case. We propose a position- and sliding-based decision guide to support clinical decision-making; its usefulness remains to be validated in larger studies.

Для цитирования: Yokoo Suguru, Okada Yukimasa, Ohno Kyotaro, Ichikawa Takahiko, Terada Chuji, Yamana Keiya. Subcapital Femoral Neck Fracture Despite Cement-Augmented Cephalomedullary Nail Fixation for an Osteoporotic Intertrochanteric Fracture: A Case Report and Position- and Sliding-Based Decision Guide.. Clinics and practice. 2025. DOI: 10.3390/clinpract16010001.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.