A novel surgical navigation technology for placement of implants in slipped capital femoral epiphysis.
СтатьяНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Oetgen Matthew E, Litrenta Jody, Koutenaei Bamshad Azizi, Cleary Kevin R
Аннотация
BACKGROUND: Fixation with a single screw is the recommended treatment for slipped capital femoral epiphysis (SCFE). Achieving optimal implant positioning can be difficult owing to the complex geometry of the proximal femur in SCFE. We assessed a novel navigation technology incorporating an inertial measurement unit to facilitate implant placement in an SCFE model.
METHODS: Guidewires were placed into 30 SCFE models, using a navigation system that displayed the surgeon's projected implant trajectory simultaneously in multiple planes. The accuracy and the precision of the system were assessed as was the time to perform the procedure.
RESULTS: Implants were placed an average of 5.3 mm from the femoral head center, with a system precision of 0.94 mm. The actual trajectory of the implant deviated from the planned trajectory by an average of 4.9° ± 2.2°. The total average procedure time was 97 seconds.
CONCLUSION: The use of computer-based navigation in a SCFE model demonstrated good accuracy and precision in terms of both implant trajectory and placement in the center of the femoral head.
METHODS: Guidewires were placed into 30 SCFE models, using a navigation system that displayed the surgeon's projected implant trajectory simultaneously in multiple planes. The accuracy and the precision of the system were assessed as was the time to perform the procedure.
RESULTS: Implants were placed an average of 5.3 mm from the femoral head center, with a system precision of 0.94 mm. The actual trajectory of the implant deviated from the planned trajectory by an average of 4.9° ± 2.2°. The total average procedure time was 97 seconds.
CONCLUSION: The use of computer-based navigation in a SCFE model demonstrated good accuracy and precision in terms of both implant trajectory and placement in the center of the femoral head.