MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1016/j.plabm.2023.e00329

Медицинская библиотека · Practical laboratory medicine · 2023

Evaluation of an automated system for the quantitation of human Herpesvirus-6 DNA from clinical specimens.

Mah Jordan, Huang ChunHong, Sahoo Malaya K, Pinsky Benjamin A

СтатьяПеревод в очереди

Аннотация

BACKGROUND

Quantitation of human herpesvirus-6 (HHV-6) DNA in clinical specimens is important for the diagnosis and management of HHV-6-associated infection and reactivation in immunocompromised patients, particularly transplant recipients.

METHODS

The analytical performance of the Altona RealStar ASR HHV-6 qPCR on the semi-automated AltoStar AM16 system was assessed using HHV-6 reference material in plasma and cerebral spinal fluid (CSF). Qualitative and quantitative agreement was determined using 123 clinical EDTA plasma specimens tested using a laboratory-developed HHV-6 qPCR.

RESULTS

The 95% Lower Limit of Detection was 20 IU/mL [95% confidence interval (CI): 10 to 29] in plasma and 78 IU/mL (95% CI: 55 to 146) in CSF. The assay was linear from 7.0 to 2.0 log10 IU/mL in both matrices. Overall agreement of the RealStar ASR HHV-6 qPCR on the AltoStar AM16 with a laboratory-developed test was 95.9% (95% CI: 90.8 to 98.7). Passing-Bablok analysis of specimens quantifiable by both methods and at levels >1000 copies/mL revealed a regression line of Y = 1.00*X-0.20, with neither systematic (95% CI Y-intercept: -0.66 to 0.26) nor proportional (95% CI slope: 0.89 to 1.10) bias compared to the reference.

CONCLUSIONS

The RealStar ASR HHV-6 qPCR on the AltoStar AM16 provides accurate quantitation for clinical monitoring of HHV-6 in immunocompromised hosts.

Для цитирования: Mah Jordan, Huang ChunHong, Sahoo Malaya K, Pinsky Benjamin A. Evaluation of an automated system for the quantitation of human Herpesvirus-6 DNA from clinical specimens.. Practical laboratory medicine. 2023. DOI: 10.1016/j.plabm.2023.e00329.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.