MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1136/jcp.48.6.556

Медицинская библиотека · Journal of clinical pathology · 1995

Histological changes associated with an artificial anterior cruciate ligament.

Barry M, Thomas S M, Rees A, Shafighian B, Mowbray M A

СтатьяПеревод в очереди

Аннотация

AIMS: To investigate the histopathological features of the synovial lining of the knee following implantation of an artificial cruciate ligament.

METHODS

Eighty two patients underwent anterior cruciate ligament reconstruction for chronic, symptomatic instability of the knee. The cruciate ligament was reconstructed with a scaffold type prosthetic ligament. All patients underwent arthroscopy at the time of cruciate reconstruction and also as a "second look" procedure at a mean 26.5 months later, at which time synovial biopsy specimens were obtained in all cases.

RESULTS

The most frequent histological finding was granulomatous synovitis in 48% (39/82) of patients. Particulate polyester debris produced a greater response than carbon fibres. In 24% (20/82), nonspecific inflammation was present and in 28% (23/82) the synovium was considered normal. When the prosthetic ligament was fully covered by soft tissue in-growth, a granulomatous reaction was present in 42% (24/57), irrespective of whether the ligament was partially or totally uncovered. Stabilisation of the joint did not prevent subsequent deterioration in the articular cartilage; other factors such as increasing patient age, interval between injury and reconstruction and altered knee kinematics after reconstruction are probably important. Granulomatous synovitis was not associated with progressive chondral changes.

CONCLUSIONS

Although present in 48% of cases, granulomatous synovitis was not shown to have adverse effects on either chondral surface or the prosthetic material of the ligament. Rupture of the implant is caused by mechanical factors and granulomatous synovitis is not responsible for implant failure.

Для цитирования: Barry M, Thomas S M, Rees A, Shafighian B, Mowbray M A. Histological changes associated with an artificial anterior cruciate ligament.. Journal of clinical pathology. 1995. DOI: 10.1136/jcp.48.6.556.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.