Acceleration of slice encoding for metal artifact correction at 0.55 T using hexagonal sampling.
Аннотация
PURPOSE: To evaluate a hexagonal sampling approach for accelerated slice encoding for metal artifact correction (SEMAC) at 0.55 T. Contemporary mid-field systems (0.1 T-1.0 T) show tremendous potential for imaging near metal implants. However, the limited parallel-imaging options necessitate alternative methods for scan time reduction.
METHODS
We apply retrospective hexagonal undersampling to current state-of-the-art SEMAC with 2-fold generalized autocalibrating partially parallel acquisitions-based parallel imaging at 0.55 T. The hexagonal sampling approach results in an additional 50% scan time reduction. Feasibility is evaluated with phantom experiments involving spinal fixation and total hip arthroplasty hardware, and in vivo experiments involving patients with spinal fusions with varying compositions and 1 volunteer with a total hip arthroplasty.
RESULTS
Hexagonal sampling provides an additional 50% scan time reduction with compatible image quality. The two tradeoffs are (i) a small increase in signal void due to the loss of signal from one SEMAC spectral bin during post-acquisition filtering and (ii) position-dependent signal-to-noise-ratio reduction at locations close to the edge of the field of view in the phase-encoding direction.
CONCLUSION
We demonstrate that hexagonal sampling can provide 50% scan time reduction in addition to generalized autocalibrating partially parallel acquisitions/parallel imaging for SEMAC at 0.55 T without introducing substantial artifacts. This may be a valuable mechanism for reducing scan time at 0.55 T and other midfield strengths, where parallel-imaging acceleration is limited.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.