DOI 10.1302/2046-3758.156.bjr-2025-0620.r1
Design and validation of a biomechanics device for preclinical arthrofibrosis models.
Аннотация
AIMS: Reliable assessment of joint stiffness is essential for studying arthrofibrosis in preclinical animal models. This study presents a new dedicated biomechanics device for measuring knee stiffness in mouse, rat, and rabbit models of arthrofibrosis, with the aim of validating its performance and comparing it to a previously validated device.
METHODS
The new system integrates a torque load cell, stepper motor, and absolute encoder in a through-hole, unified configuration for precise continuous torque-angle measurement, with custom limb brackets ensuring species-specific alignment at the joint's centre of rotation. To validate this device and compare it to a previously validated device, arthrofibrosis was induced using established models of extra-articular immobilization in mice and intra-articular violation with immobilization in rats and rabbits, followed by defined remobilization periods. Torque-angle curves were recorded ex vivo across species-specific torque ranges, and measurement variability between the new and a previous validated device was statistically compared using Brown-Forsythe's test.
RESULTS
The onset of hyperextension in contralateral limbs occurred at mean angles of 152.4° (mice), 154.4° (rats), and 152.8° (rabbits), and at maximum torque, mean flexion angles were 79.2°, 87.3°, and 140.5°, respectively. No significant differences in measurement variability were observed between devices across all species.
CONCLUSION
These findings demonstrate that our new device provides reproducible torque-angle data while maintaining the statistical robustness of the previous device and offering improved mechanical design and operational standardization. By enabling standardized, cross-species joint stiffness quantification, this platform improves cross-study comparability and strengthens the translational reliability of preclinical arthrofibrosis research.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.