MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1001/jamanetworkopen.2023.16480

Медицинская библиотека · JAMA network open · 2023

Effect of a Continuous Bedside Pressure Mapping System for Reducing Interface Pressures: A Randomized Clinical Trial.

Ho Chester, Ocampo Wrechelle, Southern Danielle A, Sola Darlene, Baylis Barry, Conly John M, Hogan David B, Kaufman Jaime, Stelfox Henry Thomas, Ghali William A

СтатьяРандомизированное исследованиеПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

IMPORTANCE

Continuous bedside pressure mapping (CBPM) technology can assist in detecting skin areas with excessive interface pressure and inform efficient patient repositioning to prevent the development of pressure injuries (PI).

OBJECTIVE

To evaluate the efficacy of CBPM technology in reducing interface pressure and the incidence of PIs.

DESIGN, SETTING, AND

PARTICIPANTS

This parallel, 2-group randomized clinical trial was performed at a tertiary acute care center. The study started to enroll participants in December 2014 and was completed in May 2018. Participants included adults partially or completely dependent for bed mobility. Statistical analysis was performed from September 2018 to December 2022.

INTERVENTION: Nursing staff using visual feedback from CBPM technology for 72 hours.

MAIN OUTCOMES AND MEASURES: Absolute number of sensing points with pressure readings greater than 40 mm Hg, mean interface pressure across all sensing points under a patient's body, proportion of participants who had pressure readings greater than 40 mm Hg, and pressure-related skin and soft tissue changes.

RESULTS

There were 678 patients recruited. After attrition, 260 allocated to the control group (151 [58.1%] male; mean [SD] age, 61.9 [18.5] years) and 247 in the intervention group (147 [59.5%] male; mean [SD] age, 63.6 [18.1] years) were included in analyses. The absolute number of sensing points with pressures greater than 40 mm Hg were 11 033 in the control group vs 9314 in the intervention group (P = .16). The mean (SD) interface pressure was 6.80 (1.63) mm Hg in the control group vs 6.62 (1.51) mm Hg in the intervention group (P = .18). The proportion of participants who had pressure readings greater than 40 mm Hg was 99.6% in both the control and intervention groups.

CONCLUSIONS AND RELEVANCE: In this randomized clinical trial to evaluate the efficacy of CBPM technology in the reduction of interface pressure and the incidence of PIs in a tertiary acute care center, no statistically significant benefit was seen for any of the primary outcomes. These results suggest that longer duration of monitoring and adequately powered studies where CBPM feedback is integrated into a multifaceted intervention to prevent PI are needed.

TRIAL REGISTRATION: ClinicalTrials.gov Identifier: NCT02325388.

Для цитирования: Ho Chester, Ocampo Wrechelle, Southern Danielle A, Sola Darlene, Baylis Barry, Conly John M, Hogan David B, Kaufman Jaime, Stelfox Henry Thomas, Ghali William A. Effect of a Continuous Bedside Pressure Mapping System for Reducing Interface Pressures: A Randomized Clinical Trial.. JAMA network open. 2023. DOI: 10.1001/jamanetworkopen.2023.16480.