A comparison of the shock-absorbing properties of cervical disc prosthesis bearing materials.

Dahl Michael C, Jacobsen Stephen, Metcalf Newton, Sasso Rick, Ching Randal P

СтатьяПеревод в очереди

Аннотация

BACKGROUND DATA: Cervical arthroplasty offers theoretical advantages over traditional spinal fusion, including elimination of adjacent segment disease and elimination of the risk of pseudoarthrosis formation. Initial studies of cervical arthroplasty have shown promising results, however, the ideal design characteristics for disc replacement constructs have not been determined. The current study seeks to quantify the differences in the shock absorption characteristics of three commonly used materials in cervical disc arthroplasty.

METHODS

Three different nucleus materials, polyurethane (PU), polyethylene (PE) and a titanium-alloy (Ti) were tested in a humidity- and temperature-controlled chamber. Ten of each nucleus type underwent three separate mechanical testing protocols to measure 1) dynamic stiffness, 2) quasi-static stiffness, 3) energy absorption, and 4) energy dissipation. The results were compared using analysis of variance.

RESULTS

PU had the lowest mean dynamic stiffness (435 ± 13 N/mm, P < .0001) and highest energy absorption (19.4 ± 0.1 N/mm, P < .0001) of all three nucleus materials tested. PU was found to have significantly higher energy dissipation (viscous damping ratio 0.017 ± 0,001, P < .0001) than the PE or TI nuclei. PU had the lowest quasi-static stiffness (598 ± 23 N/mm, P < .0001) of the nucleus materials tested. A biphasic response curve was observed for all of the PU nuclei tests.

CONCLUSIONS

Polyurethane absorbs and dissipates more energy and is less stiff than either polyethylene or titanium.

LEVEL OF EVIDENCE: Basic Science/Biomechanical Study.

CLINICAL RELEVANCE: This study characterizes important differences in biomechanical properties of materials that are currently being used for different cervical disc prostheses.

Для цитирования: Dahl Michael C, Jacobsen Stephen, Metcalf Newton, Sasso Rick, Ching Randal P. A comparison of the shock-absorbing properties of cervical disc prosthesis bearing materials.. SAS journal. 2011. DOI: 10.1016/j.esas.2011.01.002.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.