Revision for Aseptic Loosening of Highly Porous Acetabular Components in Primary Total Hip Arthroplasty: An Analysis of 20,993 Total Hip Replacements.

Hoskins Wayne, Rainbird Sophia, Holder Carl, Graves Stephen E, Bingham Roger

СтатьяПеревод в очереди

Аннотация

BACKGROUND

Highly porous-coated titanium acetabular components have a high coefficient of friction and ultraporous surfaces to enhance bone ingrowth and osseointegration in total hip arthroplasty (THA). There have been concerns with the development of early radiolucent lines and aseptic loosening of highly porous acetabular components. It is unclear whether these concerns relate to a specific implant or the entire class. The aim of this study is to compare the revision rates for aseptic loosening of highly porous acetabular combinations in primary THA using data from a large joint replacement registry.

METHODS

Data were retrieved from the Australian Orthopedic Association National Joint Replacement Registry for the study period September 1999 to December 2019. All primary THA procedures recorded and performed for osteoarthritis using the most common combinations for each highly porous acetabular component with highly cross-linked polyethylene and a 32-mm or 36-mm femoral head were included. The primary outcome measure was revision for aseptic loosening of the acetabular component. Results were adjusted for patient age and gender.

RESULTS

There were 20,993 primary THA procedures performed for osteoarthritis using a highly porous acetabular component across 6 combinations. Relative to the POLARSTEM/R3 (StikTite), the Exeter V40/Tritanium had a significantly higher risk of revision for aseptic loosening of the acetabular component (hazard ratio 0.21, 95% confidence interval 0.06-0.74, P = .014). There was no difference between any other highly porous acetabular component combination and no late revisions for aseptic loosening.

CONCLUSION

Highly porous-coated titanium acetabular components have low rates of aseptic loosening with long-term follow-up. A difference between components may exist.

LEVEL OF EVIDENCE: Level III.

Для цитирования: Hoskins Wayne, Rainbird Sophia, Holder Carl, Graves Stephen E, Bingham Roger. Revision for Aseptic Loosening of Highly Porous Acetabular Components in Primary Total Hip Arthroplasty: An Analysis of 20,993 Total Hip Replacements.. The Journal of arthroplasty. 2021. DOI: 10.1016/j.arth.2021.10.011.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.