[3D-MR myelography in diagnosis of lumbar spinal nerve root compression syndromes. Comparative study with conventional myelography].
Аннотация
65 patients with nerve root compression syndrome were examined using a new type of MR-technique, which is comparable to the conventional X-ray myelography. The results of the prospective case study were compared with previous clinical experiences (1). For the examinations a 1.0T whole body MR-system (Siemens Magnetom Impact) was used. A strong T2*-weighted 3D-FISP sequence (TR = 73 ms, TE = 21 ms, alpha = 7 degrees) was applied in sagittal orientation using a circularly polarized oval spine coil. To obtain fat suppression a frequency selective 1-3-3-1 prepulse was applied prior to the imaging sequence. The acquired 3D-data set was evaluated using a Maximum Intensity Projection (MIP) program. Our results confirmed earlier experiences which showed that the diagnostic sensitivity of 3D-MR myelography (3D-MRM) is comparable to that of conventional X-ray myelography. In cases of severe spinal canal stenosis and spondylolisthesises, and in cases of postoperative scar tissue with nerve root compressions, the sensitivity of the 3D-MRM is higher as compared to that of conventional X-ray myelography.