DOI 10.1080/08941939.2025.2520264
Comparative Analysis of Biomechanical Stability and Pain Reduction in Novel TLIF Devices.
Аннотация
OBJECTIVE
This study aims to evaluate the biomechanical and clinical performance of a new anti-backout TLIF device compared to a traditional device.
METHODS
This study involved a rat model, with biomechanical tests including static axial compression, static and dynamic settlement, and blade-cutting torque tests. Pain behavior in rats (n = 6) and material compatibility through cell toxicity and hemolysis tests were also assessed.
RESULTS
The modified anti-backout interbody fusion cage demonstrated a yield load of 7747.36 ± 274.96 N in static axial compression testing, significantly higher than the traditional TLIF cage's 6933.36 ± 65.00 N (p < 0.05), indicating superior load resistance. In static settlement testing, the modified cage's yield load was 1020.87 ± 13.22 N, also notably higher than the traditional cage's 939.06 ± 8.03 N (p < 0.05). In static pullout testing, the maximum pullout force of the modified cage with the blade extended reached 534.02 ± 21.24 N, exceeding the 476.97 ± 24.45 N without the blade (p < 0.05), showing advantages in maximum pullout force and stiffness. Biocompatibility tests revealed lower cytotoxicity and a hemolysis rate of less than 5% for the modified cage material, significantly better than the traditional material's 8% (p < 0.05).
CONCLUSION
The new anti-backout TLIF device provides enhanced stability, reduced pain, and improved material compatibility, supporting its potential for clinical application.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.