MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1007/s43390-025-01150-8

Медицинская библиотека · Spine deformity · 2025

Robotic-assisted pedicle screw placement improves accuracy, precision, and breach rates compared with freehand and image-guided navigation techniques: a cadaveric study.

Kazarian Gregory S, Groisser Benjamin N, Thakur Ankush, Dekhne Mihir S, Hillstrom Howard J, Kaidi Austin C, Mok Jung Kee, Adhiyaman Akshitha, Tracey Olivia C, Zucker Colson, Wisch Jenna L, Cunningham Matthew, Hresko M Timothy, Haddas Ram, Blanco John, Mintz Douglas N, Breighner Ryan E, Widmann Roger F, Heyer Jessica H

СтатьяСравнительное исследованиеПеревод в очереди

Аннотация

STUDY

DESIGN

Cadaveric study.

PURPOSE: Compare the accuracy and precision of robotic navigation (RAN), freehand (FH) and freehand navigation (NAV) techniques for pedicle screw placement.

METHODS

Three cadavers were assigned to the FH, NAV, and RAN methods. Cadavers were CT scanned preoperatively, then underwent bilateral T1-L5 screw planning. RAN and NAV screws were placed by one surgeon, while the FH screws were placed by a second surgeon. Automated computer vision techniques were used to assess the tip, tail, and mid-pedicle screw positions compared to the preoperative plan along three axes on postoperative CT scans. Systematic error, precision, and accuracy were defined by signed mean error, two standard deviations (SDs), and mean absolute error (MAE), respectively. A factorial ANOVA with pairwise comparisons was used with post-hoc t-tests using Bonferroni corrections.

RESULTS

Eighty-seven screws were placed in three cadavers (24 RAN, 29 NAV, 34 FH). Fifteen screws were skipped due to registration difficulties (10 RAN, 5 NAV). RAN outperformed NAV and FH in regard to MAE angular deviation from planned screw trajectory (p < 0.001), the mid-pedicle in the superior-inferior direction (p = 0.002), and at the screw tip in all planes (p < 0.05), with a maximum error of 1.29 mm and 1.98° (vs. 2.29 mm and 5.23° for NAV and 5.67 mm and 11.18° for FH). No RAN screws had medial breach (1 (3.4%) for NAV; 4 (11.8%) for FH).

CONCLUSIONS

In our study, RAN demonstrated the best accuracy and precision in the majority of measured parameters, when compared to FH and NAV techniques.

Для цитирования: Kazarian Gregory S, Groisser Benjamin N, Thakur Ankush, Dekhne Mihir S, Hillstrom Howard J, Kaidi Austin C, Mok Jung Kee, Adhiyaman Akshitha, Tracey Olivia C, Zucker Colson, Wisch Jenna L, Cunningham Matthew, Hresko M Timothy, Haddas Ram, Blanco John, Mintz Douglas N, Breighner Ryan E, Widmann Roger F, Heyer Jessica H. Robotic-assisted pedicle screw placement improves accuracy, precision, and breach rates compared with freehand and image-guided navigation techniques: a cadaveric study.. Spine deformity. 2025. DOI: 10.1007/s43390-025-01150-8.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.