DOI 10.1177/11297298261469954

Медицинская библиотека · The journal of vascular access · 2026

A novel ultrasound-based vein visualization technology can better detect optimal peripheral veins when compared to landmark.

Liddelow Michael D, Hansen Ellaby L, Jackson Gavin M, Molan Kate C, Ho Phuoc Hao, Doyle Barry J, Bappoo Nikhilesh, Carr Peter J

СтатьяПеревод в очереди

Аннотация

INTRODUCTION

Peripheral intravenous catheter (PIVC) failure prior to therapy completion is a major issue. Technology to identify more suitable veins for cannulation is essential to improve successful insertion and reduce post-insertion complications and avoidable costs.

OBJECTIVE

The experimental study was designed to compare vessel selection for cannulation when using a novel ultrasound-based vein visualization prototype versus traditional landmark technique.

METHODS

Three clinicians competent in cannulation (CCC) used landmark technique to select and mark a vein indicated for short peripheral catheter (SPC) placement on each arm of the 12 patients recruited (24 arms). This process was repeated using the vein visualization prototype. A blinded Vascular Access Specialist (VAS) evaluated each marked site using ultrasound (GE Logiq S7). Vessel cannulation suitability was rated on a 5-point Likert (⩾3 considered suitable). Ethics approval was obtained (SMHS HREC RGS0000006930).

RESULTS

Landmark-selected veins were deemed suitable less often (52.2% vs 78.3%; p = 0.08) than prototype-selected veins and had lower average suitability scores (2.57 vs 3.22; p = 0.03). Landmark technique selected veins from various sites: hand (4.3%), wrist (17.4%), mid-forearm (73.9%), and antecubital fossa (4.3%), whereas the prototype predominantly selected mid-forearm veins (95.7%). Landmark-selected veins were significantly shallower (1.83 mm vs 2.61 mm; p = 0.03) than prototype-selected veins, however the diameter of these veins were not significantly different between the two techniques (2.20 mm vs 2.72 mm; p = 0.05). Both techniques detected veins with 95.8% sensitivity.

CONCLUSION

The prototype outperformed the landmark technique by visualizing non-palpable, non-visible vessels. The veins identified were larger and optimally located in the mid-forearm, both attributes linked to fewer complications and improved first-time insertion success. Future prototype refinement will include ergonomic improvements and clinical data to enhance performance.

Для цитирования: Liddelow Michael D, Hansen Ellaby L, Jackson Gavin M, Molan Kate C, Ho Phuoc Hao, Doyle Barry J, Bappoo Nikhilesh, Carr Peter J. A novel ultrasound-based vein visualization technology can better detect optimal peripheral veins when compared to landmark.. The journal of vascular access. 2026. DOI: 10.1177/11297298261469954.