Background/

Objectives

The combined effect of femoral neck-shaft angle and lag screw position on unstable intertrochanteric fracture fixation has not been well established. This biomechanical study evaluated the effects of two caput-collum-diaphyseal (CCD) angles and two lag screw positions on construct stability.

Methods

Twenty-four synthetic femurs with identical AO/OTA 31-A2.2 fracture gaps (2 mm) were fixed using cephalomedullary nails with CCD angles of either 125° or 130°, each with a central or inferior (calcar) lag screw (n = 6/group). Constructs were tested in a single-leg stance under preloading, cyclic loading (75-750 N, 10,000 cycles, and 2 Hz), and axial loading to failure. Lag screw migration was measured radiographically, and femoral head rotation was recorded using a three-dimensional coordinate-measuring device. Stiffness, failure load, and rotations were compared using the Kruskal-Wallis and Bonferroni post hoc tests.

Results

The 125° inferior configuration showed the highest stiffness (188 ± 15 N/mm, p = 0.038) and failure load (1350 ± 97 N, p = 0.047), with the least screw migration (0.54 ± 0.11 mm, p = 0.003), significantly outperforming the 125° central and 130° central constructs. However, it exhibited greater varus collapse (2.25 ± 0.27°, p = 0.013) and axial rotation (~20-30% higher than others, p = 0.025). Screw position had a stronger effect on stability than the CCD angle, although the 130° inferior construct showed slightly less varus deformation.

Conclusions

An inferior calcar-guided lag screw improves fixation strength and stiffness in unstable intertrochanteric fractures, particularly in those with a 125° nail. However, this configuration increases varus and rotational displacement, warranting adjunct measures to enhance rotational control in clinical applications.

Для цитирования: Yoon Yong-Cheol, Lee Sung-Jae, Song Hyung Keun. Biomechanical Analysis of Fixation Strength in Unstable Intertrochanteric Femoral Fracture Models Based on the Caput-Collum-Diaphyseal Angle of Cephalomedullary Nails and Position of Lag Screws.. Journal of clinical medicine. 2025. DOI: 10.3390/jcm14186495.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.