Аннотация

PURPOSE: To compare the accuracy of pedicle screw placement in adolescent idiopathic scoliosis (AIS) between robotics and navigation and clarify the factors that cause screw deviation when robotics is used.

METHODS

Fifty consecutive patients who underwent posterior spinal fusions with computer-assisted pedicle screw placement including robotics and navigation for AIS were included. A total of 741 pedicle screws (250: Robot group, 491: Navi group) were evaluated on postoperative CT images. A rate of penetration of ≥ 2 mm was calculated as the deviation rate. After propensity score matching, we examined vertebral levels, the distance from the reference frame (RF), and the pedicle channel grade as factors for deviation.

RESULTS

The deviation rate was significantly lower in the Robot group than in the Navi group (Robot group: 1.6%, Navi group: 7.5%). After propensity score matching, 22 cases were extracted. At T5-T8, the deviation rate of the Robot group was significantly lower than that of the Navi group. In the Robot group, the T2-T4 deviation rate was significantly higher than at the other vertebral levels. The distance from the RF didn't affect the deviation rate. The deviation rate of pedicle channel Grade 4 (inner diameter of less than 1 mm) was significantly higher than for the other grades.

CONCLUSION

The deviation rate of robotics was 1.6%, lower than that of navigation. The narrow pedicles with an inner diameter of less than 1 mm (deviation rate: 22.2%) and the upper thoracic level (deviation rate: 14.3%) were factors related to screw deviation even when using robotics.

Для цитирования: Akazawa Tsutomu, Torii Yoshiaki, Ueno Jun, Umehara Tasuku, Iinuma Masahiro, Yoshida Atsuhiro, Tomochika Ken, Ohtori Seiji, Niki Hisateru. Accuracy of computer-assisted pedicle screw placement for adolescent idiopathic scoliosis: a comparison between robotics and navigation.. European spine journal : official publication of the European Spine Society, the European Spinal Deformity Society, and the European Section of the Cervical Spine Research Society. 2022. DOI: 10.1007/s00586-022-07502-6.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.