[Utility of robotic-assisted total knee arthroplasty in patients with an occupied femoral canal].
Аннотация
INTRODUCTION
the presence of implants that occupy the femoral canal is frequent in patients undergoing ipsilateral total knee replacement (TKR). The use of electronic alignment and robotic assistance make intramedullary alignment unnecessary and could be adequate in situations with an occupied femoral canal (OFC).
MATERIAL AND
METHODS
we present a prospective cohort of 25 patients who underwent robotic alignment TKR and had prior ipsilateral surgery in the femur that resulted in occupied femoral canal.
RESULTS
we had 21 female, four male patients, with an average age of 62 years. There were 14 left knees, and 11 right knees. Twenty-two patients had a hip arthroplasty and three had femur osteosynthesis. Alignment prior to surgery was 16 valgus knees (average 11°, 3-22°) and nine varus knees (average 9.5°, 0.5-18.5°). TKR alignment was five neutral (0° as measured with the robot), 11 valgus 1.3° (range 0.5-3.5°) and 8 varus 1.5° (range 0.5- 2.5°). The average preoperative WOMAC scores was 43 points, postoperative 14 points.
CONCLUSIONS
we present a cohort of patients with moderate to severe varus and valgus knee deformities with an occupied femoral canal that were electronic alignment during robot-assisted (RA) TKR without the use of an intramedullary alignment (IMA) rod. We conclude electronic alignment of TKR can be performed successfully in patients with occupied femoral canal.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.