Early mechanical failure patterns after femoral neck fixation with the femoral neck system and dynamic hip screw.

Baetz Clémence, Maupetit Adrien, Marc Clément, Rony Louis, David Guillaume

СтатьяПеревод в очереди

Аннотация

BACKGROUND

The Femoral Neck System (FNS) has been introduced as a minimally invasive alternative to the Dynamic Hip Screw (DHS) for intracapsular femoral neck fracture fixation. Comparative data on early mechanical failure patterns remain limited.

PURPOSE: To describe the early mechanical complications observed after FNS and DHS fixation during a period of progressive FNS adoption, without assuming implant-specific causality.

METHODS

This retrospective single-center cohort included 130 consecutive patients treated between December 2020 and January 2024 (42 DHS, 88 FNS). The primary outcome was mechanical revision within one year. Tip-apex distance (TAD) and postoperative reduction quality using the Garden Alignment Index (GAI) were assessed on immediate postoperative radiographs.

RESULTS

Mechanical revision occurred in 3 DHS patients (7.1%; 95% CI 1.5-19.5%) and 8 FNS patients (9.1%; 95% CI 4.0-17.1%; p = 0.25). Reduction was acceptable in 40/42 DHS and 80/88 FNS patients (p = 0.50). Mechanical failure occurred in 5/10 patients with unacceptable reduction versus 6/120 with acceptable reduction (p < 0.001). Mean TAD was similar between groups (18.8 ± 4.8 versus 19.6 ± 5.2 mm; p = 0.34), and all four cut-outs occurred with TAD >25 mm. Observed failure patterns included shortening/medialization and cut-out after DHS, and shortening/medialization, cut-out, and peri-implant fracture after FNS.

CONCLUSION

Different mechanical complications were observed after DHS and FNS fixation, but confounding by indication and the small number of events preclude implant-specific causal conclusions. Reduction quality and implant positioning appear important determinants of early mechanical failure.

Для цитирования: Baetz Clémence, Maupetit Adrien, Marc Clément, Rony Louis, David Guillaume. Early mechanical failure patterns after femoral neck fixation with the femoral neck system and dynamic hip screw.. European journal of orthopaedic surgery & traumatology : orthopedie traumatologie. 2026. DOI: 10.1007/s00590-026-04955-z.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.