DOI 10.1097/brs.0000000000005384

Медицинская библиотека · Spine · 2025

Pretapping Is Not Effective in Preventing Pedicle Fractures in Lumbar Spinal Instrumentation: A Biomechanical Cadaveric Study.

Bauer David E, Mari Major Luca, Wigger Oliver, Rohner Ramon, Farshad Mazda, Widmer Jonas, Betz Michael

СтатьяПеревод в очереди

Аннотация

STUDY

DESIGN

Biomechanical cadaveric study.

OBJECTIVE

To evaluate whether pretapping reduces the risk of pedicle fractures following pedicle screw instrumentation.

SUMMARY OF BACKGROUND DATA: Pedicle screw instrumentation is the standard for rigid fixation of the thoracolumbar spine. While complications such as screw loosening and misplacement are well-documented, pedicle fractures are rare and poorly understood. Pretapping, a technique used during screw insertion, has been proposed to enhance fixation strength and potentially reduce the risk of fractures. However, its biomechanical effects on pedicle integrity remain unclear.

METHODS

Twenty lumbar vertebrae (L1-L5) were harvested from four cadavers. Preoperative CT scans were used to plan screw trajectories and assess bone mineral density (BMD). Each vertebra was alternately instrumented with screws using pretapping and non-pretapping techniques. Biomechanical testing measured maximum bending force to failure under a flexion moment. Pedicle fractures were classified based on CT imaging, and statistical analysis was performed to evaluate factors influencing fracture risk.

RESULTS

The maximum bending force to failure showed no significant difference between pretapping (103.9±47.5 Nm) and non-pretapping (98.1±43.4 Nm) groups ( P =0.321). Pretapping significantly reduced maximum insertion torque (1.30±1.05 vs. 2.91±2.65 Nm, P =0.025). Pedicle fill was a significant predictor of bending force (β=222.29 Nm, P =0.035), while BMD and pedicle diameter were not. Fracture patterns did not differ significantly between groups ( P =0.384).

CONCLUSIONS

Pre-tapping does not significantly influence the biomechanical stability of pedicle screws or the risk of pedicle fractures under flexion moments. However, pedicle fill is a critical factor in screw fixation strength. These findings suggest that optimizing pedicle fill is more relevant than pre-tapping in enhancing construct stability.

Для цитирования: Bauer David E, Mari Major Luca, Wigger Oliver, Rohner Ramon, Farshad Mazda, Widmer Jonas, Betz Michael. Pretapping Is Not Effective in Preventing Pedicle Fractures in Lumbar Spinal Instrumentation: A Biomechanical Cadaveric Study.. Spine. 2025. DOI: 10.1097/brs.0000000000005384.