DOI 10.1016/j.artd.2024.101450

Медицинская библиотека · Arthroplasty today · 2024

Surgeon-Applied Stress and a Ligament Tensor Instrument Provide a Similar Assessment of Preresection Flexion Laxity During Robotic Total Knee Arthroplasty.

Woelfle Catelyn A, Weiner Travis R, Sculco Peter K, Sarpong Nana O, Shah Roshan P, Cooper H John

СтатьяПеревод в очереди

Аннотация

BACKGROUND

Robotic-assisted total knee arthroplasty (RA-TKA) allows surgeons to perform intraoperative soft tissue laxity assessments prior to bone resections and is used to alter resections to achieve gap balance. This study compared 2 techniques for flexion gap laxity assessment during RA-TKA.

METHODS

A prospective study of 50 primary RA-TKAs performed by a single surgeon was conducted between February and October 2023. Following full exposure, anterior tibial dislocation, and osteophyte removal, maximal medial and lateral compartment flexion laxity was quantified to the nearest 0.5 mm by the robotic system using a dynamic, surgeon-applied stress (SURGEON). This data was used to plan a balanced flexion gap by adjusting the femoral component size, rotation, and anterior-posterior translation. Flexion laxity was quantified again after distal femoral and proximal tibial resections using a ligament tensor instrument (TENSOR). These new data were used to plan for the same desired flexion gap using the same variables. Paired-samples t-tests and a simple linear regression were used for analysis.

RESULTS

Both methods produced near-identical recommendations for femoral component sizing (mean deviation 0.06 sizes, range -1 to +1 size; P = .569), rotation (deviation mean 1.0°, range -3.0° to +3.0°; P = .741), and anterior-posterior translation (deviation mean 0.13 mm, range -0.5 to +0.5 mm, P = .785). SURGEON femoral component rotation predicted TENSOR rotation (R2 = 0.157; 95% confidence interval = 0.124, 0.633; P = .004).

CONCLUSIONS

Assessing flexion laxity with a surgeon-applied stress vs a ligament tensor produced near-identical laxity data in RA-TKA, suggesting surgeons may comfortably choose either technique as a reliable method.

LEVEL OF EVIDENCE: Level III.

Для цитирования: Woelfle Catelyn A, Weiner Travis R, Sculco Peter K, Sarpong Nana O, Shah Roshan P, Cooper H John. Surgeon-Applied Stress and a Ligament Tensor Instrument Provide a Similar Assessment of Preresection Flexion Laxity During Robotic Total Knee Arthroplasty.. Arthroplasty today. 2024. DOI: 10.1016/j.artd.2024.101450.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.