DOI 10.1186/s12891-025-08654-4
Prediction of screw loosening by measuring the insertion torque in non-osteoporotic patients: an in vitro study.
Аннотация
BACKGROUND
Pedicle screws are commonly used in spinal surgeries, but screw loosening remains a major concern, even in non-osteoporotic patients. Predicting pedicle screw stability via the insertion torque is a controversial topic, mainly studied on osteoporotic cadavers. Whether the insertion torque is suitable for patients with healthy bone mineral density (BMD) remains unknown. The aim was to investigate the influencing factors, namely insertion torque, BMD, screw diameter, length, surface area, volume, screw-in rotations, vertebral level, on the screw loosening stability during distractions and to understand if intra-operative predictions are possible.
METHODS
Non-osteoporotic thoraco-lumbar vertebrae (n = 50) were used to implant five different pedicle screws (n = 100) while measuring the insertion torque. After embedding the endplates, the force needed to distract the screw head by 1 mm was tested.
RESULTS
The insertion toque (2.3 ± 0.9 Nm) showed the highest influence on the distraction force (324.8 ± 84.4 N) followed by the screw size and vertebral level. BMD did not show any effects.
CONCLUSIONS
The linear correlation of insertion torque and the bending force suggests an alternative prediction metric for screw loosening which could improve the outcome of surgeries and patients' safety. This is potentially a simple, intra-operative method, which can be used in future.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.