Clinical feasibility of camera-type X-ray for intraoperative imaging of hip arthroplasty: a pilot study.

СтатьяНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Lee Kyu-Chong, Kim Joo Hui, Hong Seok Ha, Ahn Kyung-Sik, Cho Yongwon, Kang Woo Young, Kim Baek Hyun, Shim Euddeum, Han Seung Beom, Kang Chang Ho

Аннотация

OBJECTIVES: To evaluate the clinical feasibility of a camera-type X-ray system, compared with conventional portable X-ray, for intraoperative imaging during hip arthroplasty.

MATERIALS AND METHODS: Twenty-one patients who underwent hip arthroplasty were prospectively enrolled. Conventional portable and camera-type X-ray images were obtained intraoperatively (n = 42). Three reviewers (two radiologists and one orthopedic surgeon) independently assessed image quality and diagnostic confidence using 4-point Likert scales, as well as the feasibility of assessing leg length discrepancies (LLD), calculated as the difference between the transischial line and the lesser trochanter (TI-LT). Two radiologists also measured the TI-LT interval in both modalities. Image quality and diagnostic confidence were compared using the Wilcoxon rank-sum test. The feasibility of LLD assessment was analyzed using Fisher's exact test. The agreement between camera-type and portable X-ray measurements was evaluated using intraclass correlation coefficients (ICCs).

RESULTS: Image quality scores were significantly higher for portable X-rays than for camera-type X-rays (mean ± SD, 3.65 ± 0.44 vs. 2.57 ± 0.89; p < 0.001). Diagnostic confidence was also higher with the portable X-rays (3.86 ± 0.27 vs. 2.95 ± 0.91; p < 0.001). However, the feasibility of the LLD assessment did not differ significantly between the two modalities for any reviewer (all p > 0.1). The TI-LT interval measurements demonstrated good agreement between camera-type and portable X-rays, with an ICC of 0.750 (95% CI 0.454-0.898).

CONCLUSIONS: Although conventional portable X-rays provided superior image quality and diagnostic confidence, camera-type X-rays demonstrated comparable feasibility for intraoperative evaluation of LLD and good agreement in quantitative measurements. These findings support the clinical feasibility of using camera-type X-rays for intraoperative imaging during hip arthroplasty.