Biomechanical Effects of Proximal Polyetheretherketone Rod Extension on the Upper Instrumented and Adjacent Levels in a Human Long-Segment Construct: A Cadaveric Model.

Pereira Bernardo de Andrada, Lehrman Jennifer N, Sawa Anna G U, Wangsawatwong Piyanat, Godzik Jakub, Xu David S, Turner Jay D, Kelly Brian P, Uribe Juan S

СтатьяПеревод в очереди

Аннотация

OBJECTIVE

The high mechanical stress zone at the sudden transition from a rigid to flexible region is involved in proximal junctional kyphosis (PJK) physiopathology. We evaluated the biomechanical performance of polyetheretherketone (PEEK) rods used as a nontraditional long semirigid transition phase from a long-segment metallic rod construct to the nonfused thoracic spine.

METHODS

Pure moment range of motion (ROM) tests (7.5 Nm) were performed on 7 cadaveric spine segments followed by compression (200 N). Specimens were tested in the following conditions: (1) intact; (2) T10-pelvis pedicle screws and rods (PSRs); and (3) extending the proximal construct to T6 using PEEK rods (PSR+PEEK). T10-11 rod strain, T9 anterolateral bone strain, and T10 screw bending moments were analyzed.

RESULTS

At the upper instrumented vertebra (UIV)+1, PSR+PEEK versus PSR significantly decreased ROM in flexion (115%, p = 0.02), extension (104%, p = 0.003), left lateral bending (46%, p = 0.02), and right lateral bending (63%, p = 0.008). Also, at UIV+1, PSR+PEEK versus intact significantly decreased ROM in flexion (111%, p = 0.01) and extension (105%, p = 0.003). The UIV+1 anterior column bone strain was significantly reduced with PSR+PEEK versus PSR during right lateral bending (p = 0.02). Rod strain polarities reversed with PEEK rods in all loading directions except compression.

CONCLUSION

Extending a long-segment construct using PEEK rods caused a decrease in adjacent-level hypermobility as a consequence of long-segment immobilization and also redistributed the strain on the UIV and adjacent levels, which might contribute to PJK physiopathology. Further studies are necessary to observe the clinical outcomes of this technique.

Для цитирования: Pereira Bernardo de Andrada, Lehrman Jennifer N, Sawa Anna G U, Wangsawatwong Piyanat, Godzik Jakub, Xu David S, Turner Jay D, Kelly Brian P, Uribe Juan S. Biomechanical Effects of Proximal Polyetheretherketone Rod Extension on the Upper Instrumented and Adjacent Levels in a Human Long-Segment Construct: A Cadaveric Model.. Neurospine. 2022. DOI: 10.14245/ns.2244146.073.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.