BACKGROUND

Total ankle arthroplasty (TAA) has continued to increase in popularity as a treatment for ankle arthritis. Despite popularity, continual updates to implant designs strive for improved outcomes which has led to different implant design changes including an extended-length tibial tray to offer more distal tibia cortical coverage. The purpose of this study was to compare radiographic findings (tibial component subsidence, tibial component loosening, heterotopic ossification [HO]) and patient-reported outcome measures (PROMs) between 2 age-, gender-, and body mass index-matched TAA groups.

METHODS

This study was a retrospective review of TAA patients matched into 2 groups: group 1 patients received the extended tibial tray implant, and group 2 received the standard tibial tray implant. Pre- and postoperative lateral radiographs and PROMs (Veterans-RAND 12 [VR-12] Item Health Survey, Ankle Osteoarthritis Scale [AOS], visual analog scale [VAS], and patient satisfaction) were analyzed.

RESULTS

Seventy-eight patients were allocated to each group. The mean age was 66.7 years in group 1 and 66.9 years in group 2, with a mean follow-up of 44.6 months and 50.7 months, respectively. There were no significant differences between the groups for pre- and postoperative PROMs. Group 1 had higher patient satisfaction with overall care (95.4 vs 89.8, P = .019). Radiographically, the rate of HO formation was significantly lower in group 1 (38.5% vs 62.8%, P < .001), and group 1 also had greater plantarflexion (P = .096) and significantly less dorsiflexion (P = .008).

CONCLUSION

Results appear to indicate that using TAA implants with an extended-length tibial tray is associated with more complete distal tibia cortical coverage, a numerically lower but statistically nonsignificant rate of tibial component subsidence, less heterotopic ossification formation, a trend toward greater postoperative plantarflexion, and higher patient satisfaction with overall care.

LEVEL OF EVIDENCE: Level III, retrospective cohort study.

Для цитирования: Traynor Christopher J, Zhang Hui, Dock Carissa C, Stone McGaver Rebecca, Coetzee J Chris, Seybold Jeffrey D, Seiffert Kayla J, Cammack Paul M, Den Hartog Bryan D. Use of an Extended-Length Tibial Tray in Total Ankle Arthroplasty: Rate of Subsidence, Loosening, Heterotopic Bone Formation, and Clinical Outcomes.. Foot & ankle orthopaedics. 2026. DOI: 10.1177/24730114261452667.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.