MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1007/s00586-025-09128-w

Медицинская библиотека · European spine journal : official publication of the European Spine Society, the European Spinal Deformity Society, and the European Section of the Cervical Spine Research Society · 2025

The Total Disc Replacement Osteolysis Grading Scale - a simple, reliable and quantifiable tool for assessing, managing and reporting osteolysis after cervical total disc replacement.

Celenza Alana, Gaff Jessica, Pabbruwe Moreica, Amaya Juan, Amlani Ashik, Berg Andrew, Häckel Sonja, Kern Michael, Miles Andrew, Taylor Paul, Cunningham Gregory

СтатьяПеревод в очереди

Аннотация

PURPOSE: Cervical Total Disc Replacement (CTDR) related periprosthetic osteolysis is inconsistently reported. The purpose of this study is to assess the previously published TDR Osteolysis Grading Scale, a quantitative classification system of CTDR-related osteolysis, and to assess its reliability and utility using Computer Tomography (CT) and X-ray imaging.

METHODS

Participants were assigned to Groups A (CT, 20 participants, 27 implants) and B (X-ray, 20 participants, 26 implants). Four blinded raters independently measured osteolytic cysts and both vertebral endplates, calculated the percentage of osteolysis and assigned osteolysis grades: Grade 0, no osteolysis; Grade 1, < 50% of 1 endplate; Grade 2, < 50% of both endplates; Grade 3, ≥ 50% of 1 endplate; Grade 4, ≥ 50% of both endplates. Inter-rater reliability and sensitivity were assessed using intraclass correlation coefficients (ICC), Fleiss' and Cohen's Kappa statistics and sensitivity testing.

RESULTS

Group A had good reliability for osteolytic cyst measurements (ICC = 0.78-0.79), poor reliability for endplate measurements (ICC = 0.41-0.45) and moderate reliability for osteolysis grades (κ-coefficient = 0.42). Group B had moderate reliability for osteolytic cyst (ICC = 0.55-0.57) and endplate measurements (ICC = 0.55-0.64) and fair reliability for osteolysis grades (κ-coefficient = 0.30). Reliability between reference grades and assigned grades was moderate for Group A (κ-coefficient = 0.46-0.60) and fair for Group B (κ-coefficient = 0.08-0.35). Sensitivity was greater for high grade osteolysis (grade 3 and 4) in Group A (0.80-1) than Group B (0.17-0.50).

CONCLUSION

The TDR Osteolysis Grading Scale demonstrated good reliability and sensitivity amongst raters with CT. Implementing this tool may aid in standardising osteolysis reporting, monitoring osteolysis progression and clinical decision-making.

Для цитирования: Celenza Alana, Gaff Jessica, Pabbruwe Moreica, Amaya Juan, Amlani Ashik, Berg Andrew, Häckel Sonja, Kern Michael, Miles Andrew, Taylor Paul, Cunningham Gregory. The Total Disc Replacement Osteolysis Grading Scale - a simple, reliable and quantifiable tool for assessing, managing and reporting osteolysis after cervical total disc replacement.. European spine journal : official publication of the European Spine Society, the European Spinal Deformity Society, and the European Section of the Cervical Spine Research Society. 2025. DOI: 10.1007/s00586-025-09128-w.