MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1007/s00586-025-08724-0

Медицинская библиотека · European spine journal : official publication of the European Spine Society, the European Spinal Deformity Society, and the European Section of the Cervical Spine Research Society · 2025

Spinous process resistance to different materials and looping techniques for interspinous lumbar vertebropexy.

Widmer Jonas, Calek Anna-Katharina, Fasser Marie-Rosa, Suter Mauro, Allen Brian, Farshad Mazda

СтатьяПеревод в очереди

Аннотация

PURPOSE: Semi-rigid spinal stabilization has been explored as an alternative to spinal fusion, and early experience with spinal segment augementation ("vertebropexy") is promising. The main technical challenge is to maintain the integrity of the spinous processes during loading. This study aimed to compare different looping materials and techniques with respect to their performance in maintaining spinous process integrity.

METHODS

One hundred and five thoracolumbar vertebrae were tested with various looping materials and techniques; the "Tunnel-only" double-loop technique was tested with a synthetic tape (FiberTape®), bovine tendon, and a hybrid tape option. Additionally, the performance of the synthetic tape was tested for other augmentation techniques such as the "Figure-of-eight" looping technique, a double-loop combination of tunneling and cortical wrapping, and a double-loop with "Cortical wrapping only". Biomechanical testing was performed by uniaxial caudo-cranial distraction to failure.

RESULTS

The loads required to cause spinous process failure were lowest with the synthetic tape, followed by tendon and hybrid constructs by tendency (419 N vs. 487 N vs. 519 N) in the "Tunnel-only" double-loop technique. The comparison showed that the "Tunnel + cortical wrapping" technique required significantly higher forces to induce failure compared to other techniques, particularly the "Tunnel only" method (p < 0.001).

CONCLUSION

The choice of the looping technique and material in lumbar interspinous vertebropexy significantly affects the resistance of the spinous process to load. Techniques that incorporate cortical bone and use tendinous material demonstrate superior resistance to higher forces, compared to methods that involve passing synthetic tape through a hole solely within trabecular bone. Additionally, the role of trabecular bone density in the spinous process is relatively minor when cortical bone is utilized as an abutment for the loop.

Для цитирования: Widmer Jonas, Calek Anna-Katharina, Fasser Marie-Rosa, Suter Mauro, Allen Brian, Farshad Mazda. Spinous process resistance to different materials and looping techniques for interspinous lumbar vertebropexy.. European spine journal : official publication of the European Spine Society, the European Spinal Deformity Society, and the European Section of the Cervical Spine Research Society. 2025. DOI: 10.1007/s00586-025-08724-0.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.