Biomechanical Impact of Cementation Technique Variations on Femoral Stem Stability: An In Vitro Polyurethane Model Study.
СтатьяНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Fazakas Roland, Bondar Laura Ioana, Toth Csongor, Osser Brigitte, Ilia Iosif, Miuta Caius Calin, Fruja Dan, Iovanovici Diana Carina, Gavrila-Ardelean Liviu, Pop Alexandru
Аннотация
Background/Objectives: Achieving optimal primary stability in cemented total hip arthroplasty remains a critical factor influencing long-term implant success. Variability in cementation techniques can significantly affect biomechanical performance, yet consensus on best practices is lacking. This study investigates the influence of cementation parameters on femoral stem fixation. Methods: This in vitro comparative study evaluated four cementation techniques-Classic (line-to-line), Press-Fit (undersized reaming), Overreaming (oversized reaming), and Valgus Malpositioning (15° deviation). An experimental model using standardized Polyurethane (PU) bone surrogates was developed. Mechanical testing assessed axial deformation and ultimate load capacity to failure. Results: The Press-Fit technique demonstrated significantly greater deformation (17.10 ± 0.89 mm) but a reduced load capacity (6317.47 ± 518.34 N) compared to the Classic approach. Overreaming and Valgus techniques both showed reduced mechanical performance, with Overreaming yielding the lowest structural integrity. Conclusions: Cement mantle thickness emerged as the primary determinant of biomechanical stability, surpassing the impact of implant positioning. While increased mantle thickness improves energy absorption, it may compromise ultimate strength. These findings underscore the importance of optimizing the cementation technique to balance flexibility and mechanical resistance, guiding surgical protocols toward improved implant longevity. This study introduces a novel integrative approach combining fluoroscopic assessment of cement mantle morphology with mechanical testing in a standardized model, providing new evidence on the relative influence of mantle thickness and implant malposition on femoral stem stability.