Intraoperative mechanical measurement of bone quality with the DensiProbe.

Hoppe Sven, Uhlmann Michael, Schwyn Robert, Suhm Norbert, Benneker Lorin M

СтатьяОбзорПеревод в очереди

Аннотация

Reduced bone stock can result in fractures that mostly occur in the spine, distal radius, and proximal femur. In case of operative treatment, osteoporosis is associated with an increased failure rate. To estimate implant anchorage, mechanical methods seem to be promising to measure bone strength intraoperatively. It has been shown that the mechanical peak torque correlates with the local bone mineral density and screw failure load in hip, hindfoot, humerus, and spine in vitro. One device to measure mechanical peak torque is the DensiProbe (AO Research Institute, Davos, Switzerland). The device has shown its effectiveness in mechanical peak torque measurement in mechanical testing setups for the use in hip, hindfoot, and spine. In all studies, the correlation of mechanical torque measurement and local bone mineral density and screw failure load could be shown. It allows the surgeon to judge local bone strength intraoperatively directly at the region of interest and gives valuable information if additional augmentation is needed. We summarize methods of this new technique, its advantages and limitations, and give an overview of actual and possible future applications.

Для цитирования: Hoppe Sven, Uhlmann Michael, Schwyn Robert, Suhm Norbert, Benneker Lorin M. Intraoperative mechanical measurement of bone quality with the DensiProbe.. Journal of clinical densitometry : the official journal of the International Society for Clinical Densitometry. 2014. DOI: 10.1016/j.jocd.2014.06.002.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.