Lumbosacral spinal compression device with the use of a cushion back support in supine MRI.

Sungkarat Witaya, Laothamatas Jiraporn, Worapruekjaru Ladawan, Hooncharoen Boonthida, Charoensuk Jarruwat, Chousangsuntorn Khaisang

СтатьяПеревод в очереди

Аннотация

BACKGROUND

We hypothesized that axial-loaded magnetic resonance imaging (MRI), modified with the use of a cushion placed behind the lower back (i.e. BS-MRI method), would simulate the standing position more accurately than an axial-loaded MRI without a cushion back support (BS).

PURPOSE: To determine whether the BS-MRI method demonstrated similar morphologies on intervertebral disc (IVD), dural sac, and spinal curvature as those detected on 90° standing MRIs in individuals with suspected spinal stenosis.

MATERIAL AND

METHODS

Twenty-five subjects underwent a BS-MRI, as well as axial-loaded and standing MRI studies. Outcome measures were four radiographic parameters of the lumbar spine: IVD height (DH); dural sac cross-sectional area (DCSA); and spinal curvature (i.e. lumbar lordosis [LL] and L1-L3-L5 angle [LA]).

RESULTS

Major differences (>5%) between standing MRI and BS-MRI methods were observed in DCSA, DH, and LL. Major differences between standing and axial loaded MRIs were observed only in DCSA and LA. Although BS-MRIs demonstrate an image of the lumbar spine curvature (i.e. LA) which is closer to that when standing than axial-loaded MRIs, it is likely to overestimate both narrowing of dural sac and extent of LL.

CONCLUSION

Using a compression device with a BS to simulate weight-bearing on the lumbar spine is not recommended due to: (i) overestimation of the narrowing of the dural sac and extent of LL; and (ii) underestimation of loss of disc height. Supine axial-loading produced DCSA and DH which were strongly correlated with those detected with standing MRIs. Exceptions were that LL and LA were underestimated.

Для цитирования: Sungkarat Witaya, Laothamatas Jiraporn, Worapruekjaru Ladawan, Hooncharoen Boonthida, Charoensuk Jarruwat, Chousangsuntorn Khaisang. Lumbosacral spinal compression device with the use of a cushion back support in supine MRI.. Acta radiologica (Stockholm, Sweden : 1987). 2020. DOI: 10.1177/0284185120951963.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.