STUDY

DESIGN

Retrospective comparative cohort study.

OBJECTIVES

Identify whether there are patient or surgical risk factors to predict the probability of failure of iliac screw fixation after correction of neuromuscular scoliosis.

SUMMARY OF BACKGROUND DATA: There are high reported failure rates of pelvic fixation in long posterior spinal fusion (PSF) constructs to the sacrum for neuromuscular scoliosis.

METHODS

Patients aged 5 to 25 years, at a single institution, had PSF to the sacrum from 2001 to 2009 with pelvic fixation using iliac screws. Clinical data were retrospectively reviewed to identify patient and surgical variables related to surgery. Failure of iliac fixation was identified strictly as a broken screw, disengagement of the screw from the connector or the connector from the rod, or set plug failure. Lucency around the screws greater than 2 mm was recorded but not considered a failure. Variables were analyzed in a statistical model to identify predictors of failure.

RESULTS

A total of 108 patients met inclusion criteria; 100 (38 female and 62 male) had appropriate radiographs and minimum 2-year follow-up (average, 5.5 years). Coronal deformity correction averaged 59%. Most patients (89%) had fill of 6 of 8 possible distal fixation points composed of L4, L5, S1, and ilium bilaterally. Iliac screw failure occurred in 27 patients (27%). The initial single predictor statistical model identified 2 possible predictors of failure (patient: spastic tone; and surgical: absence of distal crosslink). In the multivariable model, spastic tone remained a predictor of failure (p = .0103), whereas absence of distal crosslink bordered on significance (p = .0516).

CONCLUSIONS

Iliac screw fixation failure is common in patients with long PSF constructs for neuromuscular scoliosis. Spastic tone is a risk factor for failure of pelvic fixation. A distal crosslink may protect against pelvic fixation failure. Alternative techniques for pelvic fixation should be studied in an attempt to improve the failure rates of iliac screw pelvic fixation.

Для цитирования: Garg Sumeet, Holland Courtney, LaGreca Jaren, McNair Bryan, Erickson Mark. Predicting Failure of Iliac Fixation in Neuromuscular Spine Deformity.. Spine deformity. 2014. DOI: 10.1016/j.jspd.2014.01.002.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.