Optimization of 3D-printed titanium interbody cage design. Part 1: in vitro biomechanical study of subsidence.

Farber S Harrison, Oldham Alton J, O'Neill Luke K, Sawa Anna G U, Ratliff Alexis C, Doomi Ahmed, Pereira Bernardo de Andrada, Uribe Juan S, Kelly Brian P, Turner Jay D

СтатьяПеревод в очереди

Аннотация

BACKGROUND CONTEXT: Cage subsidence is a complication of interbody fusion associated with poor clinical outcomes. 3D-printed titanium interbody cages allow for the alteration of features such as stiffness and porosity. However, the influence of these features on subsidence and their biological effects on fusion have not been rigorously evaluated.

PURPOSE: This 2-part study sought to determine how changes in 3D-printed titanium cage parameters affect subsidence using an in vitro bone model (Part 1) and biological fusion using an in vivo sheep model (Part 2).

STUDY

DESIGN

Biomechanical foam block model.

METHODS

In Part 1 of this study, 9 implant types were tested (8 per implant type). The implant types included 7 3D-printed titanium interbody cages with various surface areas, porosities, and surface topographies, along with 1 standard polyetheretherketone (PEEK) cage and 1 solid titanium cage. Subsidence testing was performed in a standardized foam block model using 2 different densities of foam. Digital imaging correlation was used to determine the relative vertical displacement of the interbody cage-foam block construct.

RESULTS

Subsidence decreased as the surface contact area with the bone model increased (all p≤.01). Increased porous surface topography increased subsidence, while a nonporous surface significantly decreased subsidence (all p<.001). Subsidence did not differ based on changes in implant porosity (all p≥.35) or material property/modulus (all p≥.19). Subsidence was significantly decreased with the higher density foam (p<.001). The stiffness of the implant was affected by porosity (all p<.02) and smooth surface topography (p=.01) but not by lumen size (all p≥.15). Stiffness did not differ between porous titanium and PEEK implants (p=.96), which were both less stiff than solid titanium implants (both p<.001). Surface area negatively correlated with subsidence (r=-0.786, p=.012) but was not correlated with stiffness (r=0.560, p=.12).

CONCLUSIONS

Implant surface area and surface topography greatly influenced interbody subsidence. Apparent stiffness, implant porosity, and material property did not affect subsidence in this in vitro model. Higher foam density also led to lower subsidence than low-density foam. Biological response in the in vivo setting likely also influences clinical subsidence, which is evaluated in the companion study (Part 2).

CLINICAL SIGNIFICANCE: This study provides valuable information regarding the new 3D-printed titanium technology. We showed that cage surface area and surface topography were the implant design parameters that had the greatest influence on the development of interbody subsidence. Moreover, bone mineral density was the factor that had the greatest effect on subsidence prevention. These data support patient optimization before surgery and emphasize the importance of endplate protection during surgery.

Для цитирования: Farber S Harrison, Oldham Alton J, O'Neill Luke K, Sawa Anna G U, Ratliff Alexis C, Doomi Ahmed, Pereira Bernardo de Andrada, Uribe Juan S, Kelly Brian P, Turner Jay D. Optimization of 3D-printed titanium interbody cage design. Part 1: in vitro biomechanical study of subsidence.. The spine journal : official journal of the North American Spine Society. 2024. DOI: 10.1016/j.spinee.2024.12.002.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.