MedElevator Войти в исследования
← Поиск публикаций

DOI 10.3233/bme-191053

Медицинская библиотека · Bio-medical materials and engineering · 2019

Breakage of metal hip arthroplasty components: Retrieval and structural analysis.

Łapaj Łukasz, Woźniak Waldemar, Wiśniewski Tomasz, Rozwalka Justyna, Paczesny Łukasz, Zabrzyński Jan, Janusz Piotr, Kruczyński Jacek

СтатьяПеревод в очереди

Аннотация

BACKGROUND

Breakage of joint arthroplasty components are rare, yet during an implant retrieval program we found several cases.

OBJECTIVE

In this study we examined the components to determine the causes and mechanisms of breakage of these implants.

METHODS

From our collection of 849 retrievals we selected 682 cases with metal parts (503 hip, 79 knee arthroplasties) and identified fractured components: seven hip resurfacing implants, five total hip replacement stems, one monopolar femoral head, and one modular revision femoral stem from. Implants were examined using optical and scanning electron microscopy; metallographic sections were prepared and samples of periprosthetic tissues underwent microscopic examination.

RESULTS

In the resurfacing components breakage occurred in small stems placed in the femoral neck due to necrosis of femoral heads, with no metal flaws detected. Fatigue breakage of femoral stems was caused by presence of material flaws in the CoCrMo alloy, and corrosion. The monopolar head failed in fatigue fracture mechanism, breakage was initiated in an undercut near the taper connection for femoral component. The modular stem from Ti alloy sustained fatigue fracture induced by corrosion caused by debris from previously revised stem; no material flaws were detected in this sample. In most cases periprosthetic tissues had a morphology typical for aseptic loosening.

CONCLUSIONS

In our series failure was caused by material flaws, presence of stress raisers and localized corrosion. Our findings indicate that sharp edges and other features which can act as stress raisers should be avoided in newly designed implants. Corrosion induced fracture of the modular Ti stem indicates the need for a detailed debridement of periprosthetic tissues during revision arthroplasties.

Для цитирования: Łapaj Łukasz, Woźniak Waldemar, Wiśniewski Tomasz, Rozwalka Justyna, Paczesny Łukasz, Zabrzyński Jan, Janusz Piotr, Kruczyński Jacek. Breakage of metal hip arthroplasty components: Retrieval and structural analysis.. Bio-medical materials and engineering. 2019. DOI: 10.3233/bme-191053.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.