MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1016/j.ard.2025.05.010

Медицинская библиотека · Annals of the rheumatic diseases · 2025

Exploring the limit of image resolution for human expert classification of vascular ultrasound images in giant cell arteritis and healthy subjects: the GCA-US-AI project.

Bauer Claus-Juergen, Chrysidis Stavros, Dejaco Christian, Koster Matthew J, Kohler Minna J, Monti Sara, Schmidt Wolfgang A, Mukhtyar Chetan B, Karakostas Pantelis, Milchert Marcin, Ponte Cristina, Duftner Christina, de Miguel Eugenio, Hocevar Alojzija, Iagnocco Annamaria, Terslev Lene, Døhn Uffe Møller, Nielsen Berit Dalsgaard, Juche Aaron, Seitz Luca, Keller Kresten Krarup, Karalilova Rositsa, Daikeler Thomas, Mackie Sarah Louise, Torralba Karina, van der Geest Kornelis S M, Boumans Dennis, Bosch Philipp, Tomelleri Alessandro, Aschwanden Markus, Kermani Tanaz A, Diamantopoulos Andreas, Fredberg Ulrich, Inanc Nevsun, Petzinna Simon M, Albarqouni Shadi, Behning Charlotte, Schäfer Valentin Sebastian

СтатьяПеревод в очереди

Аннотация

OBJECTIVES

Prompt diagnosis of giant cell arteritis (GCA) with ultrasound is crucial for preventing severe ocular and other complications, yet expertise in ultrasound performance is scarce. The development of an artificial intelligence (AI)-based assistant that facilitates ultrasound image classification and helps to diagnose GCA early promises to close the existing gap. In the projection of the planned AI, this study investigates the minimum image resolution required for human experts to reliably classify ultrasound images of arteries commonly affected by GCA for the presence or absence of GCA.

METHODS

Thirty-one international experts in GCA ultrasonography participated in a web-based exercise. They were asked to classify 10 ultrasound images for each of 5 vascular segments as GCA, normal, or not able to classify. The following segments were assessed: (1) superficial common temporal artery, (2) its frontal and (3) parietal branches (all in transverse view), (4) axillary artery in transverse view, and 5) axillary artery in longitudinal view. Identical images were shown at different resolutions, namely 32 × 32, 64 × 64, 128 × 128, 224 × 224, and 512 × 512 pixels, thereby resulting in a total of 250 images to be classified by every study participant.

RESULTS

Classification performance improved with increasing resolution up to a threshold, plateauing at 224 × 224 pixels. At 224 × 224 pixels, the overall classification sensitivity was 0.767 (95% CI, 0.737-0.796), and specificity was 0.862 (95% CI, 0.831-0.888).

CONCLUSIONS

A resolution of 224 × 224 pixels ensures reliable human expert classification and aligns with the input requirements of many common AI-based architectures. Thus, the results of this study substantially guide projected AI development.

Для цитирования: Bauer Claus-Juergen, Chrysidis Stavros, Dejaco Christian, Koster Matthew J, Kohler Minna J, Monti Sara, Schmidt Wolfgang A, Mukhtyar Chetan B, Karakostas Pantelis, Milchert Marcin, Ponte Cristina, Duftner Christina, de Miguel Eugenio, Hocevar Alojzija, Iagnocco Annamaria, Terslev Lene, Døhn Uffe Møller, Nielsen Berit Dalsgaard, Juche Aaron, Seitz Luca, Keller Kresten Krarup, Karalilova Rositsa, Daikeler Thomas, Mackie Sarah Louise, Torralba Karina, van der Geest Kornelis S M, Boumans Dennis, Bosch Philipp, Tomelleri Alessandro, Aschwanden Markus, Kermani Tanaz A, Diamantopoulos Andreas, Fredberg Ulrich, Inanc Nevsun, Petzinna Simon M, Albarqouni Shadi, Behning Charlotte, Schäfer Valentin Sebastian. Exploring the limit of image resolution for human expert classification of vascular ultrasound images in giant cell arteritis and healthy subjects: the GCA-US-AI project.. Annals of the rheumatic diseases. 2025. DOI: 10.1016/j.ard.2025.05.010.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.