DOI 10.1002/acr.25350

Медицинская библиотека · Arthritis care & research · 2024

Diagnostic Accuracy of Raman Spectroscopy Integrated With Polarized Light Microscopy for Calcium Pyrophosphate-Associated Arthritis.

Niessink Tom, Janssen Matthijs, Giesen Tanja, Efdé Monique N, Comarniceanu Antoaneta C, Otto Cees, Jansen Tim L

СтатьяПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

OBJECTIVE

We studied the performance of integrated Raman polarized light microscopy (iRPolM) for the identification of calcium pyrophosphate (CPP)-associated arthritis (CPPD).

METHODS

This is a diagnostic accuracy study including 400 consecutive synovial fluid samples from a single hospital in the Netherlands. Accuracy measures were calculated against polarized light microscopy (PLM) and the 2023 American College of Rheumatology (ACR)/EULAR criteria set for CPPD.

RESULTS

The interrater reliability between iRPolM and the 2023 ACR/EULAR criteria set for CPPD was strong (κ = 0.88). The diagnostic performance of iRPolM compared to the 2023 ACR/EULAR criteria set was sensitivity 86.0% (95% confidence interval [CI] 73.3-94.2), specificity 99.1% (95% CI 97.5-99.8), positive likelihood ratio 100.33 (95% CI 32.3-311.3), negative likelihood ratio 0.14 (95% CI 0.07-0.28), positive predictive value 93.5% (95% CI 82.2-97.8), negative predictive value 98.0% (95% CI 82.2-97.8), and accuracy 97.5% (95% CI 95.5-98.8). We allowed rheumatologists to rate the certainty of their microscopic identification of CPP and found a large correspondence between iRPolM and a certain identification (κ = 0.87), whereas only 10% of the uncertain CPP identifications could be confirmed with iRPolM. We identified several novel particle types in synovial fluid analysis, including calcium carbonate crystals, deposited carotenoids, microplastics, and three types of Maltese cross birefringent objects.

CONCLUSION

iRPolM can easily identify CPP crystals with a strong diagnostic performance. PLM alone is not specific enough to reliably resolve complicated cases, and the implementation of Raman spectroscopy in rheumatology practice can be of benefit to patients with suspected CPPD.

Для цитирования: Niessink Tom, Janssen Matthijs, Giesen Tanja, Efdé Monique N, Comarniceanu Antoaneta C, Otto Cees, Jansen Tim L. Diagnostic Accuracy of Raman Spectroscopy Integrated With Polarized Light Microscopy for Calcium Pyrophosphate-Associated Arthritis.. Arthritis care & research. 2024. DOI: 10.1002/acr.25350.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.