MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1016/j.spinee.2024.09.008

Медицинская библиотека · The spine journal : official journal of the North American Spine Society · 2024

Stability of medially and laterally malpositioned screws: a biomechanical study on cadavers.

Tsagkaris Christos, Fasser Marie-Rosa, Farshad Mazda, Passaplan Caroline, Cornaz Frederic, Widmer Jonas, Spirig José Miguel

СтатьяПеревод в очереди

Аннотация

BACKGROUND CONTEXT: Pedicle screw instrumentation is widely used in spine surgery. Axial screw misplacement is a common complication. In addition to the recognized neurovascular risks associated with screw misplacement, the biomechanical stability of misplaced screws remains a subject of debate.

PURPOSE: The present study investigates whether screw misplacement in the lumbar spine reduces mechanical screw hold.

STUDY DESIGN/

SETTING

Cadaveric biomechanical study.

METHODS

Pedicle screw (mis)placement was planned for 12 fresh frozen cadaveric spines between the T12 and the L5 levels. The screws were then implanted into the vertebrae with the help of 3D-printed template guides. Pre- and postinstrumentation computed tomography (CT) scans were acquired for instrumentation planning and quantification of the misplacement. The instrumented vertebrae were potted into CT transparent boxes using Polymethyl methacrylate and mounted on a standardized biomechanical setup for pull-out (PO) testing with uniaxial tensile load.

RESULTS

The bone density of all the specimens as per HU was comparable. The predicted pull-out force (POF) for screws medially misplaced by 2 , 4, and 6 mm was respectively 985 N (SD 474), 968 N (SD 476) and 822 N (SD 478). For screws laterally misplaced by 2 , 4, and 6 mm the POF was respectively 605 N (SD 473), 411 N (SD 475), and 334 N (SD 477). Screws that did not perforate the pedicle (control) resisted pull-out forces of 837 N (SD 471).

CONCLUSIONS

Medial misplacement is associated with increased axial screw hold against static loads compared to correctly placed screws and laterally placed screws.

CLINICAL SIGNIFICANCE: In clinical settings, the reinsertion of medially misplaced screws should primarily aim to prevent neurological complications while the reinsertion of lateral misplaced screws should aim to prevent screw loosening.

Для цитирования: Tsagkaris Christos, Fasser Marie-Rosa, Farshad Mazda, Passaplan Caroline, Cornaz Frederic, Widmer Jonas, Spirig José Miguel. Stability of medially and laterally malpositioned screws: a biomechanical study on cadavers.. The spine journal : official journal of the North American Spine Society. 2024. DOI: 10.1016/j.spinee.2024.09.008.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.