Mechanical Fatigue of Titanium Dental Implants After Implantoplasty: An In Vitro Study Combined with Finite Element Simulations.

Padullés-Roig Esteban, Sevilla Pablo, Velasco-Ortega Eugenio, Cerrolaza Miguel, Fonseca Darcio, Parache Jeanne, Aparicio Conrado, Gil Javier

СтатьяПеревод в очереди

Аннотация

The increasing prevalence of peri-implantitis has led to a growing clinical use of implantoplasty, a procedure involving intraoral machining of the dental implant surface to remove biofilm. The absence of standardized clinical protocols may contribute to premature fatigue failure of dental implants. The present study aimed to evaluate the influence of machining depth on the cyclic mechanical behavior of dental implants. A total of 250 commercially pure grade 4 titanium dental implants were distributed into four groups according to machining depth: untreated (original), 0.2 mm, 0.4 mm, and 0.6 mm wall reduction. The implant system featured an internal connection with a thread height of 0.4 mm. Finite element analysis was performed for each machining depth to evaluate von Mises stress distribution and simulate fatigue behavior. The numerical models were validated through experimental fatigue testing using a servo-hydraulic MTS Bionix testing machine under ISO 14801:2016 conditions, showing a high correlation between simulated and experimental results (correlation coefficients > 0.9). The results indicated that maximum von Mises stresses were concentrated at the junction between the implant thread and the implant body. The fatigue limit of the untreated implants was approximately 351 N. Implants subjected to 0.4 mm machining exhibited a fatigue limit of 301 N, whereas lower fatigue limits were observed for 0.2 mm (255 N) and 0.6 mm (185 N) reductions. These findings suggest a significant mechanical effect of thread removal: 0.4 mm implantoplasty may provide improved fatigue performance compared to 0.2 mm, potentially due to reduced stress concentration at the thread-body junction. At high applied loads, fracture occurred in the coronal region of the implant, whereas at lower loads failure shifted to the implant-abutment connection. Although a good agreement between numerical and experimental results was observed, these findings should be interpreted with caution due to the in vitro testing conditions and the assumptions inherent to the finite element simulations. Therefore, while the results suggest that implantoplasty depth should not exceed the original thread height, further validation under clinically relevant conditions is required to confirm its impact on long-term mechanical reliability.

Для цитирования: Padullés-Roig Esteban, Sevilla Pablo, Velasco-Ortega Eugenio, Cerrolaza Miguel, Fonseca Darcio, Parache Jeanne, Aparicio Conrado, Gil Javier. Mechanical Fatigue of Titanium Dental Implants After Implantoplasty: An In Vitro Study Combined with Finite Element Simulations.. Journal of functional biomaterials. 2026. DOI: 10.3390/jfb17050221.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.