BACKGROUND

Recently, interspinous stabilization with the interspinous process device (IPD) has become an alternative to treat lumbar spinal stenosis. The biomechanical influence of different design features of IPDs on intradiscal pressure (IDP) and facet joint force (FJF) has not been fully understood. The aim of this study was to investigate the biomechanical performance of different IPDs using finite element (FE) method.

METHODS

A FE model of the L1-5 segments was developed and validated. Four surgical FE models were constructed by inserting different implants at the L3-4 segment (Coflex-F, DIAM, Wallis, and pedicle screw system). The 4 motion modes were simulated.

RESULTS

The IPDs decreased range of motion (ROM) at the surgical level substantially in flexion and extension, but little influence was found in lateral bending and torsion. Compared with the DIAM and Wallis devices, the Coflex-F device showed advantages in stabilizing the surgical level, especially in flexion and extension, while it increased FJF at adjacent levels by 26%-27% in extension. Among the 3 IPDs, the DIAM device exhibited the most comparable ROM, IDP, and FJF at adjacent levels compared with the intact lumbar spine. The influence of the Wallis device was between that of the Coflex-F and DIAM devices.

CONCLUSIONS

Compared with rigid fixation, the IPDs demonstrated less compensation at adjacent levels in terms of ROM, IDP, and FJF, which may lower the incidence of adjacent segment degeneration in the long term.

Для цитирования: Shen Hangkai, Fogel Guy R, Zhu Jia, Liao Zhenhua, Liu Weiqiang. Biomechanical Analysis of Different Lumbar Interspinous Process Devices: A Finite Element Study.. World neurosurgery. 2019. DOI: 10.1016/j.wneu.2019.04.051.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.