DOI 10.4103/ijvsm.ijvsm_6_26

Медицинская библиотека · International journal of veterinary science and medicine · 2026

Enhancing Accuracy and Efficiency of Canine Synovial Fluid Analysis with Hyaluronidase Pretreatment and Automated Analyzer.

Dewi Winda Kusuma, Radtanakatikanon Araya, Akaraphutiporn Ekkapol, Nedumpun Teerawut, Wangdee Chalika

СтатьяПеревод в очереди

Аннотация

BACKGROUND

Synovial fluid examination is vital for diagnosing joint diseases in dogs. However, conventional manual methods are labor-intensive and prone to imprecision. Pretreating synovial fluid with hyaluronidase and utilizing an automated analyzer may enhance the accuracy, precision, and efficiency of synovial fluid analysis, particularly in dogs.

AIMS AND

OBJECTIVES

This study aimed to assess the accuracy, precision, and efficiency of automated analyzer compared with manual cell count (MC) for analyzing synovial fluid in dogs.

MATERIALS AND

METHODS

Fifty-two synovial fluid samples with varying cellularity were evaluated for the total nucleated cell count (TNCC) and differential cell count using three methods: hyaluronidase-treated automated cell count (AC-Hy), hyaluronidase-treated MC (MC-Hy), and untreated MC. The accuracy and precision of each method were statistically compared.

RESULTS

AC-Hy and MC-Hy provide the strongest agreement in determining TNCC (mean difference: 23.92%, 95% confidence interval: -53.23%-101.07%), with AC-Hy exhibiting the highest precision (coefficient of variations [CV]: 4.42%-4.49%). For differential counts, AC-Hy performed well in high TNCC samples (CV: 0.74%-2.25%) but was less reliable in low TNCC samples. MC-Hy, however, demonstrated consistent accuracy in determining differential count across all TNCC levels and produced superior smear quality compared with MC. In addition, AC-Hy offers shorter analytical times and is technically more efficient.

CONCLUSION

In conclusion, pretreating canine synovial fluid with hyaluronidase enhances the accuracy and precision of synovial fluid analysis. While an automated analyzer enables rapid TNCC measurement, its reliability for differential counts decreases at low TNCC. Integrating automated count with hyaluronidase-treated manual differential counts and clinical evaluations would enhance diagnostic precision.

Для цитирования: Dewi Winda Kusuma, Radtanakatikanon Araya, Akaraphutiporn Ekkapol, Nedumpun Teerawut, Wangdee Chalika. Enhancing Accuracy and Efficiency of Canine Synovial Fluid Analysis with Hyaluronidase Pretreatment and Automated Analyzer.. International journal of veterinary science and medicine. 2026. DOI: 10.4103/ijvsm.ijvsm_6_26.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.