MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1016/j.wneu.2017.04.031

Медицинская библиотека · World neurosurgery · 2017

Comparative Study Between Cobalt Chrome and Titanium Alloy Rods for Multilevel Spinal Fusion: Proximal Junctional Kyphosis More Frequently Occurred in Patients Having Cobalt Chrome Rods.

Han Sanghyun, Hyun Seung-Jae, Kim Ki-Jeong, Jahng Tae-Ahn, Kim Hyun-Jib

Сравнительное исследованиеОценочное исследованиеСтатьяПеревод в очереди

Аннотация

BACKGROUND

The use of titanium alloy (Ti) rods is frequently associated with rod fracture after spinal fixation. To address this issue, cobalt chrome (CoCr) rods, which are advantageous because of their greater strength and resistance to fatigue relative to Ti rods, have been introduced. The purpose of the present study was to compare radiographic outcomes after the use of Ti versus CoCr rods in a matched cohort of patients undergoing posterior spinal fusion for treatment of spinal instability.

METHODS

We retrospectively reviewed data from patients who had undergone spinal fusion involving more than 3 levels at a single institution between 2004 and 2015. Patients were matched for age, diagnosis, 3-column osteotomy, levels fused, and T score. Fifty patients with Ti rods were identified and appropriately matched to 50 consecutive patients with CoCr rods.

RESULTS

The distributions of age at surgery, sex, diagnosis, 3-column osteotomy, levels fused, number of patients with previous surgical procedures, and T score did not significantly differ between the 2 groups. However, there were significant differences in length of follow-up (CoCr, 25.0 vs. Ti, 28.5 months; P < 0.001), fusion rate (CoCr, 45 [90%] vs. Ti, 33 [66%]; P = 0.004), occurrence of rod breakage (CoCr, 0 vs. T, 8 [16%]; P = 0.006), and junctional kyphosis (CoCr, 24 [46%] vs. Ti, 9 [18%]; P = 0.003).

CONCLUSIONS

Our findings indicate that the use of CoCr rods is effective in ensuring stability of the posterior spinal construct and accomplishment of spinal fusion. Furthermore, our results indicate that junctional kyphosis may occur more frequently in CoCr systems than in Ti systems.

Для цитирования: Han Sanghyun, Hyun Seung-Jae, Kim Ki-Jeong, Jahng Tae-Ahn, Kim Hyun-Jib. Comparative Study Between Cobalt Chrome and Titanium Alloy Rods for Multilevel Spinal Fusion: Proximal Junctional Kyphosis More Frequently Occurred in Patients Having Cobalt Chrome Rods.. World neurosurgery. 2017. DOI: 10.1016/j.wneu.2017.04.031.