Surveillance of Dual-Mobility Hip Systems: Damage Mode and Clinical Data Analysis.
СтатьяПоддержка исследования: вне правительства СШАНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Ahlgren Conner D, Beydoun Rami S, Maxwell Michael J, Cascardo Camilla A, Karadsheh Mark S, Moore Drew D
Аннотация
BACKGROUND: Hip instability following total hip arthroplasty (THA) is among the most common indications for revision surgery. The implantation of dual-mobility (DM) systems, designed to improve stability, continues to rise, and thus, characterizing in vivo implant damage modes is paramount.
METHODS: Under an implant retrieval protocol, 51 DM THA systems were analyzed. Each component was examined for macroscopic damage and microscopically graded for fretting corrosion at the trunnion interface. Clinically relevant data were collected from medical records.
RESULTS: Revision indications included mechanical complications (n = 15, 27%), infection (n = 12, 22%), and dislocation (n = 9, 16%). The average duration of implantation was 12 months, which significantly correlated with summed femoral head taper corrosion (P = 0.044). Average summed fretting and corrosion scores were 2.9 and 3.0 for heads (two regions, 2 to 8 summed scores possible) and 7.6 and 7.1 for trunnions (four quadrants, 4 to 16 summed scores possible), respectively. Scratching (n = 26, 65.4%) was the most common damage mode on articular surfaces of acetabular cups, burnishing (n = 29, 55.2%) on metal liners, and edge deformation (n = 22, 45.5%) on polyethylene liners. Screw-liner corrosion was noted on 10.3% (n = 3) of available acetabular liners; an increased incidence of acetabular and femoral osteolysis was noted when this damage mode was present (P = 0.019 and P = 0.022, respectively).
CONCLUSIONS: This series demonstrated in vivo damage of DM THA components following short-term-to-midterm (zero to five years) implantation, with overall mild-to-moderate fretting and corrosion scores. Femoral head taper corrosion correlated with longer implantation, and amplified trunnion corrosion was observed in the setting of infection. Screw-liner corrosion may be related to acetabular and femoral osteolysis. This study indicates trends that merit further evaluation.
METHODS: Under an implant retrieval protocol, 51 DM THA systems were analyzed. Each component was examined for macroscopic damage and microscopically graded for fretting corrosion at the trunnion interface. Clinically relevant data were collected from medical records.
RESULTS: Revision indications included mechanical complications (n = 15, 27%), infection (n = 12, 22%), and dislocation (n = 9, 16%). The average duration of implantation was 12 months, which significantly correlated with summed femoral head taper corrosion (P = 0.044). Average summed fretting and corrosion scores were 2.9 and 3.0 for heads (two regions, 2 to 8 summed scores possible) and 7.6 and 7.1 for trunnions (four quadrants, 4 to 16 summed scores possible), respectively. Scratching (n = 26, 65.4%) was the most common damage mode on articular surfaces of acetabular cups, burnishing (n = 29, 55.2%) on metal liners, and edge deformation (n = 22, 45.5%) on polyethylene liners. Screw-liner corrosion was noted on 10.3% (n = 3) of available acetabular liners; an increased incidence of acetabular and femoral osteolysis was noted when this damage mode was present (P = 0.019 and P = 0.022, respectively).
CONCLUSIONS: This series demonstrated in vivo damage of DM THA components following short-term-to-midterm (zero to five years) implantation, with overall mild-to-moderate fretting and corrosion scores. Femoral head taper corrosion correlated with longer implantation, and amplified trunnion corrosion was observed in the setting of infection. Screw-liner corrosion may be related to acetabular and femoral osteolysis. This study indicates trends that merit further evaluation.