Expert review of medical devices · 2024
Advancing robot-guided techniques in lumbar spine surgery: a systematic review and meta-analysis.
СтатьяМетаанализСистематический обзорНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Komolafe Temitope Emmanuel, Zhou Liang, Zhao Wenlong, Guo Jiachen, Li Zongdao, Fan Zhiping, Komolafe Blessing Funmi, Wei Wang, Samuel Oluwarotimi Williams
Аннотация
BACKGROUND: Lumbar spine surgery is a crucial intervention for addressing spinal injuries or conditions affecting the spine, often involving lumbar fusion through pedicle screw (PS) insertion. The precision of PS placement is pivotal in orthopedic surgery. This systematic review compares the accuracy of robot-guided (RG) surgery with free-hand fluoroscopy-guided (FFG), free-hand without fluoroscopy-guided (FHG), and computed tomography image-guided (CTG) techniques for PS insertion.
METHODS: A systematic search of various databases from 1 January 2013 to 30 December 2023 was conducted following PRISMA guidelines. Primary outcomes, including PS insertion accuracy and breach rate, were analyzed using a random-effects model. Risk of bias was assessed using the Newcastle-Ottawa Scale.
RESULTS: The overall accuracy of PS insertion using RG, based on 37 studies involving 3,837 patients and 22,117 PS, is 97.9%, with a breach rate of 0.021. RG demonstrated superior accuracy compared to FHG and CTG, with breach rates of 3.4 and 0.015 respectively for RG versus FHG, and 3.8 and 0.026 for RG versus CTG. Additionally, RG was associated with reduced mean estimated blood loss compared to CTG, indicating improved safety.
CONCLUSIONS: The RG is associated with enhanced accuracy of PS insertion and reduced breach rates over other methods. However, additional randomized controlled trials comparing these modalities are needed for further validation.
PROSPERO REGISTRATION: CRD42023483997.
METHODS: A systematic search of various databases from 1 January 2013 to 30 December 2023 was conducted following PRISMA guidelines. Primary outcomes, including PS insertion accuracy and breach rate, were analyzed using a random-effects model. Risk of bias was assessed using the Newcastle-Ottawa Scale.
RESULTS: The overall accuracy of PS insertion using RG, based on 37 studies involving 3,837 patients and 22,117 PS, is 97.9%, with a breach rate of 0.021. RG demonstrated superior accuracy compared to FHG and CTG, with breach rates of 3.4 and 0.015 respectively for RG versus FHG, and 3.8 and 0.026 for RG versus CTG. Additionally, RG was associated with reduced mean estimated blood loss compared to CTG, indicating improved safety.
CONCLUSIONS: The RG is associated with enhanced accuracy of PS insertion and reduced breach rates over other methods. However, additional randomized controlled trials comparing these modalities are needed for further validation.
PROSPERO REGISTRATION: CRD42023483997.